11. Maintenance
11.1 Regular inspection
Perform following check-ups regularly or depending on the situation.
11.2 Troubleshooting
If a trouble is recognized, check following items and perform troubleshooting.
11.2.1 Measuring value not displayed or displayed lower
Inspection items
Countermeasures
1) Check that power is supplied to the thermometer
and power supply voltage is in the normal range.
Supply normal power supply voltage.
2) Check that temperature of measuring surface is
not out of thermometer's measuring range.
Use a thermometer corresponding to the measuring
range.
3) Check that setting of emissivity in the main unit
is not too high.
Refer to 8.1 Emissivity setting and set correct emissivity.
4) Check that optical path is not interfered.
Refer to 4.3 Measuring Diameter and Measuring
Distance and assure the field of view.
11.2.2 Measuring value displayed higher
Inspection items
Countermeasures
1) Check that temperature of measuring surface is
not exceeding thermometer's measuring range.
Use a thermometer corresponding to the measuring range.
2) Check that setting of emissivity is not too low.
Refer to 8.1 Emissivity setting and set correct emissivity.
11.2.3 Indication fluctuates
Inspection items
Countermeasures
1) Check that power supply voltage is in normal
range.
Supply normal power supply voltage.
2) Check that there is no obstacle such as vapor in
the optical path.
Remove the vapor by air purge etc.
3) Check for temperature fluctuation on the
measuring surface.
Refer to 8.1 Emissivity setting, 8.3 Signal modulation
selection, and 8.4 Modulation degree setting and set
emissivity and signal modulation again.
4) Check for emissivity of measuring surface for
change
11.2.4 oFL or uFL is displayed
Indication
Cause
Countermeasures
oFL
Temperature is exceeding maximum
temperature of measuring range.
Refer to 11.2.2 Measuring value displayed higher
uFL
Temperature is below minimum
temperature of measuring
range
.
Refer to 11.2.1 Measuring value not displayed or
displayed lower