INST.No.INE-559-P1
-22-
・
Default setting is no (no reset)
【
Peak hold (no reset)
】
Sub display
T
R
:
reset time (0.0 to 99.9sec)
After peak (valley) temperature
detection, the temperature is held
during T
R
.
If setting is 0.0, hold time is less
than 0.1 sec.
If using external rest, following 9.7
Selection
of
contact
input
of
H.rSt(hold
reset)
is
required.
External reset is reset at external
contact input start up.
(55msec or more is necessary for
contact input start up detecting
pulse width)
【
Peak hold (internal reset)
】
【
Peak hold (external reset)
】
External
contact input
9.4.1 Selection of hold reset type
(Register this function if peak or valley hold is selected at previous section.)
1)After register of "PEAK" or "VALL" at the previous section,
press the SEL key to display "H.rSt" shown on the right
figure on sub display.
2) Press
↑
key or
↓
key and change hold reset types.
Selection is completed by changing.
3) It shifts to the next screen by pressing the SEL key.
[Reset method types]
no
: no reset
in
: internal reset
Et
: external reset
CONT
PEAK
AL
A H
Tb
MEM
CONT
PEAK
AL
A H
Tb
MEM
[Internal reset]
CONT
PEAK
AL
A H
Tb
MEM
[External reset]
Main display
[No reset]
R e m a r k s
Caution
Caution
Input
Output
T
R
T
R
T
R
Input
Output
T
R
Input
Output