INST.No.INE-559-P1
-7-
5.2 Connections to input and output terminals
5.2.1 Terminal arrangement and terminal nameplate
Connecting cable IR-ZBP
(Color)
(Marker)
FG
Grounding
Gray
FG
8-36VDC
Power supply
Red
8-36VDC
0VDC
White/Red
0VDC
SA
Communication
Black
SA
SB
White/Black
SB
DI
Contact input
Yellow
DI
DO
Contact output
White/Yellow
DO
COM
Common
White/Brown
COM
A GND
Analog ground
White/Green
A GND
OUT-A mA
Analog output
Green
OUT-A/TC
OUT-TC
Thermocouple
output
OUT-HT
Detector
temperature output
Brown
OUT-HT
※
Analog output and TC output can not be used at the same time. Please connect either signal to
be used
.
5.2.2 Connection of power supply terminal
5.2.3 Connection of grounding terminal
5.2.4 Connection to receiving devices
Analog output is non-isolated 0 or 4 to 20mA DC, thermocouple EMF equivalent
voltage.
Connect output terminal and receiving devices to each other.
Contact output is open collector output.
Use connecting external device rated (30V, 50mA) or lower.
Use 8 to 36VDC for power supply
Applicable power wire is 0.2 to 0.5mm
2
.
Provide a low impedance earth ground (grounding resistance 100Ω or lower)
connection.
W a rn i ng
Caution
Caution
R e m a r k s
!
When closing the lid after connection, confirm the packing is not sticking out from
the groove.
Caution
※
Make sure to turn OFF the power supply for preventing an electric shock when
connecting power terminal.
Summary of Contents for IR-BZ Series
Page 2: ......
Page 11: ...INST No INE 559 P1 5 4 3 Measuring Diameter and Measuring Distance Unit mm φ16 φ8 400 200 ...