INST.No.INE-559-P1
-20-
9.3.2 Analog dummy output2
(This is valid only if "tCJ", "tCk", "tCr" or "tCS" is selected for analog output selection)
1) Press the SEL key couple of times from engineering mode
screen to display
"Out2"
shown on the right figure on sub
display.
2) Press
↑
key or
↓
key and change set value (temperature) of
analog dummy output2.
Setting is completed by changing the numeric value.
3) It shifts to the next screen by pressing the SEL key.
・
Setting range is -50 to 3500 (°C)
Output
EMF
corresponding
to
set
temperature abd selected thermocouple
Output range is as below.
tCJ: -50 to 1200°C
tCk: -50 to 1370°C
tCr: -50 to 1760°C
tCS: -50 to 1760°C
However, outputtable max. temperature
range is from -50 to 1100°C and upper
limit depends on the unit from 1030 to
1100°C.
・
Default setting is 0 (°C)
[Analog dummy output2]
CONT
PEAK
AL
A H
Tb
MEM
Sub display
Main display
R e m a r k s