INST.No.INE-559-P1
-4-
4. Installation
4.1 Installation of Main unit
Fix the Main unit by two installation holes.
4.2 Installation of Detector
Detector is fixed by screw. Make a hole on desired equipment and fix by attachment nut.
Fix a cable without looseness.
6
2
Do not use the thermometer at following places where;
●Dust, particles and corrosive gas are in the atmosphere
●Electric noise and static electricity are generated frequently
●Ambient temperature is
65°C or higher or -10°C or lower
●Ambient
temperature changes widely or humid is high
●Close to high frequency line or power line
●
Mechanical vibration and impact exist
●
Flammable or explosive gases exist
Unit: mm
(12)
80
(17)
72
7
0
1
9
.5
30
Cable gland
for power supply and
signal lines
Installation hole
2-
φ
4.5
Cable gland
for Detector (not removal)
10
Compatible cable
diameter
:φ
6.5 to
φ
8
6
2
Unit: mm
28
10
M12 P1
φ
1
4
Cable (1.5, 3, 8 or 15m)
Outer diameter:
φ
3.3
Caution
Cable vibration will be the cause of noise so please fix it to not to vibrate.
Cable allowable bending radius is 20mm.
When screwing the nut, do not co-around the detector. If co-around then it will
cause cable disconnection.
Caution