INST.No.INE-559-P1
-13-
【dELy】
【PEAk】
【VALL】
8.3 Signal modulation mode selection
Select form to modulate measurement signal (original signal).
Select "dELy" to display REAL that is based on original signal
and set modulation degree to 0.0.
1) Press the SEL key couple of times from measurement mode
screen to display "Modu" shown on the right figure on sub
marker.
2) Press
↑
key or
↓
key and change signal modulation.
Selection is completed by changing the display.
3) It shifts to the next screen by pressing the SEL key.
[
Modes of signal modulation
]
dELy
It displays the value that is based on first order lag of
time constant set at 8.4.1 Setting of modulation time
constant. When signal value fluctuates, it displays in
smooth form
PEAK
If temperature is rising, it displays the value based on
original signal.
If temperature is dropping, the value is displayed based
on attenuation rate setting. (Refer to 8.4.2 Attenuation/
Incarease rate setting)
VALL
VALL is the opposite process of PEAK
.
If temperature is dropping, the value is displayed based
on original signal.
If temperature is rising, it displays the value is displayed
based
on
increase
rate
setting.
(Refer
to
8.4.2
Attenuation/Increase rate setting)
・
Default setting is "dELy"
CONT
PEAK
AL
A H
Tb
MEM
[PEAK]
Main display
CONT
PEAK
AL
A H
Tb
MEM
Sub display
[dELy]
CONT
PEAK
AL
A H
Tb
MEM
[VALL]
R e m a r k s
Display
Origina
l
Display
Original
Display
ispl
Original