12
Instruction Manual
SMART-LIFT Electric Ceiling Lifts
CONNECT THE LIFT TO THE POWER SUPPLY
WARNING: Plenum rating requires hardwiring through the
provided strain relief, use of Class II control wiring, and all
open holes in housing (except joist tabs) must be sealed using
foil tape. See wiring instructions in this manual and follow all
local building and electrical codes.
NOTE: If the lift is cycled up and down repeatedly, the ther-
mal overload on the motor will stop operation. Operation will
resume when the thermal overload resets (3 to 5 minutes).
Non Plenum Rated Installation
1.
Plug the lift’s power cord (supplied) into the electrical out-
let in the ceiling plenum.
Plenum Rated Installation
WARNING: A licensed electrician should
disconnect and terminate the leads to the
power cord receptacle and, using the
access holes provided for a strain relief,
hard wire the unit to a power source.
WARNING: Failure to disconnect and terminate power leads
properly may result in equipment damage or personal injury.
1.
Disconnect and terminate the electrical leads from the
power cord receptacle at the interior junction box (see
Figure 45).
2.
Remove the electrical knockout and install the strain
relief (provided).
2.
Hardwire the unit to a 110V 60 Hz or 230V 50 Hz (as
appropriate) 15-amp power source.
Terminate Power Leads
Electrical Junction Box
inside Smartlift
Figure 12. Disconnect and Terminate
Power Cord Leads
Summary of Contents for SL-100
Page 20: ...19 Instruction Manual SMART LIFT Electric Ceiling Lifts THIS PAGE INTENTIONALLY BLANK ...
Page 22: ...21 Instruction Manual SMART LIFT Electric Ceiling Lifts THIS PAGE INTENTIONALLY BLANK ...
Page 24: ...23 Instruction Manual SMART LIFT Electric Ceiling Lifts THIS PAGE INTENTIONALLY BLANK ...
Page 26: ...25 Instruction Manual SMART LIFT Electric Ceiling Lifts THIS PAGE INTENTIONALLY BLANK ...
Page 29: ...Instruction Manual SMART LIFT Electric Ceiling Lifts 28 This Page Intentionally Blank ...