Intelect® Transport Combo
50
OPERATION
STARTING, STOPPING, AND INTERRUPTING ULTRASOUND THERAPY CONTINUED
NOTE:
The green light located on the back of the applicator flashes intermittently
when the sound head breaks contact with the patient.
Also, when the applicator becomes uncoupled, the message "Uncoupled"
displays
in
the
status
window.
16.
Press the Up and Down Arrow buttons to select the manner in which you want to be
notified
when
the
sound
head
becomes
uncoupled.
The
following
list
details
these
options:
• Pause timer & beep - the timer stops and the unit beeps once.
• Pause timer, no beep - the timer stops but the unit does not give an audible tone.
• Run timer & beep - the timer continues to count down and the unit beeps once.
• Run timer, no beep - the timer continues to count down, but the unit does not give
an
audible
tone.
NOTE:
When the applicator becomes uncoupled during treatment, it continues to
distribute
ultrasound
energy.
17.
Press the Enter button to accept the selection.
Do not apply the Ultrasound Applicator to the patient during the Head
Warming period. Applicator must remain in Applicator Hook duing the
Head Warming period.