Intelect® Transport Combo
SPECIFICATIONS
29
ELECTROMAGNETIC COMPATIBILITY TABLES
Guidance and manufacturer’s declaration – electromagnetic emissions
The Intelect Transport Combo is intended for use in the electromagnetic environment specified below. The
customer or the user of the Intelect Transport Combo should assure that it is used in such an environment.
Emission tests
Compliance
Electromagnetic environment -
guidance
RF emissions
CISPR 11
Group 1
The Intelect Transport Combo uses
RF energy only for its internal
function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class A
The Intelect Transport Combo is
suitable for use in all
establishments, including
domestic establishments and
those directly connected to the
public low-voltage power supply
network that supplies buildings
used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations
IEC 61000-3-3
Complies
Guidance and manufacturer’s declaration – electromagnetic immunity
The Intelect Transport Combo is intended for use in the electromagnetic environment specified below. The
customer or the user of the Intelect Transport Combo should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic
environment - guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6kV contact
±8kV air
±6kV contact
±8kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2kV for power supply
lines
±1kV for input/output
lines
±2kV for power supply
lines
±1kV for input/output
lines
Mains power quality
should be that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
±1kV differential mode
±2kV common mode
±1kV differential mode
±2kV common mode
Mains power quality
should be that of a typical
commercial or hospital
environment.
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
) for 0.5
cycle
40% U
T
(60% dip in U
T
) for 5
cycles
70% U
T
(30% dip in U
T
) for 25
cycles
<5% U
T
(>95% dip in U
T
) for 5 sec
<5% U
T
(>95% dip in U
T
) for 0.5
cycle
40% U
T
(60% dip in U
T
) for 5
cycles
70% U
T
(30% dip in U
T
) for 25
cycles
<5% U
T
(>95% dip in U
T
) for 5 sec
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user
of the Intelect Transport
Combo requires
continued operation
during power mains
interruptions, it is
recommended that the
Intelect Transport Combo
be powered from an
uninterrupted power
supply or a battery.
Power frequency
(50/60Hz) magnetic field
IEC 61000-4-8
3 V/m
3 V/m
Power frequency
magnetic fields should be
at levels characteristic of
a typical location in a
typical commercial or
hospital environment.
NOTE U
T
is the a.c mains voltage prior to application of the test level.