MVE Biological Systems
TEC 3000 Operating and Maintenance Freezer Manual
13289499 Rev H •
71
7.2.5.4. One Fill All Fill (OFAF)
This section describes how to adjust the various OFAF settings. For more information
and instructions on how to setup an OFAF network, refer to the OFAF Network Setup
section of this manual.
NOTE: Security Level 3 or higher is required to adjust the OFAF settings.
1. Press “SETUP”
Controller will prompt for a password. A flashing cursor
on the entry screen will make it clear which digit is being
changed. Use the “▲/▼” keys to scroll to the
appropriate number. Press “ENTER” to advance the
cursor to the next position.
2. Press “ENTER”
The display will read “Temperature Menus”.
3. Press “SETUP”
Until the display reads
“Advanced Settings”.
4. Press “ENTER”
The display will read “Advanced Fill Menus”.
5. Press “SETUP”
Until the display reads “OFAF Menus”.
6. Press “ENTER”
The display will read “OFAF ID”. Use the “▲/▼” keys to
adjust the OFAF ID and press “ENTER” to save.
7. Press “SETUP”
The display will read “OFAF Units”. Use the “▲/▼” keys
to adjust the number of units and press “ENTER” to
save.
8. Press “SETUP”
The display will read “One Fill All Fill”. Use the “▲/▼”
keys to adjust the type and press “ENTER” to save.
Higher User Level
Required use ▲▼ to
ENTER Password
0000
Press ENTER for
Temperature Menus
or press SETUP for
next menu
Press ENTER for
Advanced Settings
or press SETUP for
next menu
Press ENTER for
Advanced Fill menus
or press SETUP for
next menu
Press ENTER for
OFAF menus
or press SETUP for
next menu
OFAF ID
1
Use ▲▼ to adjust
Press ENTER to save
OFAF Units
3
Use ▲▼ to adjust
Press ENTER to save
One Fill All Fill
SEQUENTIAL
Use ▲▼ to adjust
Press ENTER to save