MVE Biological Systems
TEC 3000 Operating and Maintenance Freezer Manual
13289499 Rev H •
61
7.2.4.3. Alarm Buzzer
The audible alarm buzzer can be disabled. This will not disable the alarms, only the
audible buzzer.
Alarms will still be displayed visually on the screen if the alarm buzzer is
inactive.
NOTE: Security Level 2 or higher is required to enable/disable the alarm buzzer.
1. Press “SETUP”
Controller will prompt for a password. A flashing cursor
on the entry screen will make it clear which digit is being
changed. Use the “▲/▼” keys to scroll to the
appropriate number. Press “ENTER” to advance the
cursor to the next position.
2. Press “ENTER”
The display will read “Temperature Menus”.
3. Press “SETUP”
Until
the display reads “Display and Output”.
4. Press “ENTER”
The display will read “Temperature Units”.
5. Press “SETUP”
Until the display reads “Advanced Display and Output”.
6. Press “ENTER”
The display will read “Alarm Buzzer”. Use the “▲/▼”
keys to
adjust and press “ENTER” to save.
Higher User Level
Required use ▲▼ to
ENTER Password
0000
Press ENTER for
Temperature Menus
or press SETUP for
next menu
Press ENTER for
Display and Output
or press SETUP for
next menu
Temperature Units
°C
Use ▲▼ to adjust
Press ENTER to save
Press ENTER for
Advanced Display
and Output or press
SETUP for next menu
Alarm Buzzer
ENABLED
Use ▲▼ to adjust
Press ENTER to save