MVE Biological Systems
TEC 3000 Operating and Maintenance Freezer Manual
13289499 Rev H •
137
Figure 18 - Mounting Location for the HEco 1800 Series
3.3.1.3. Remove nuts from studs.
3.3.1.4. Place the battery backup on the bolts with the washers between the battery backup and the
platform. Position the battery backup with the pigtail facing towards the front of the freezer. Note: The
battery backup is capable of being positioned with the pigtail facing forwards or backwards. It is
recommended that the pigtail faces forward so the batteries inside the housing remain in an upright
position.
3.3.1.5. Fasten nuts to studs. This will secure the battery backup to the workstation on the freezer.
3.3.1.6. Please see Section 3.4 to complete the installation.
3.4. Verify that the TEC 3000 Controller will power on using the AC Power Supply only.
NOTE:
Do
not connect the battery backup to the main wire harness until the AC Power Supply has been verified
to power up the controller.
3.5. Connect the battery backup to the proper connector on the TEC 3000 Wiring Harness. See
Figure 8.
Figure 15 - Battery Backup Connector. [Black
– From the Battery Backup.
Dark Blue/Orange
– From the TEC 3000 Wiring Harness]
3.6. Please reference the Additional Feature Settings Section to view and adjust the TEC 3000
settings and to enable the battery backup feature.