MVE Biological Systems
TEC 3000 Operating and Maintenance Freezer Manual
13289499 Rev H •
126
3. Remove the four Phillips screws holding the backing plate.
4. Cut the plastic zip ties and remove the batteries noting the wire connection. Batteries
are connected in series, - to + as shown in the below diagram.
5. Ensure the
6. Reverse the removal procedure
cables are routed as shown:
Negative Terminal