MVE Biological Systems
TEC 3000 Operating and Maintenance Freezer Manual
13289499 Rev H •
150
TEC 3000 ASCII Command List
Control Commands
*IDN?
Identification Query
Input: *IDN?
Returned:
MVE TEC 3000, Software ver. X.XX
Comments:
Used to query the controller ID. Command must include “*”
preface. Returns controller model and firmware version.
ALMS
Set Alarm Status
Input: ALMS 0 or ALMS 1
Returned:
No response
Comments:
Used to set the audible alarm status where 0 = Off and 1 =
On. Has same function as the Alarm Mute key.
ALMS?
Alarm Status Query
Input: ALMS?
Returned:
0 or 1
Comments:
Used to query the current audible alarm status where 0 =
Off and 1 = On.
CALTA
Calibrate Temp A in LN2 (Single Point)
Input: CALTA
Returned:
No response
Comments:
Used to perform a single point calibration of Temp A.
Temp A sensor must be submerged in LN2. See the Temperature
Calibration section of this manual.
CALTB
Calibrate Temp B in LN2 (Single Point)
Input: CALTB
Returned:
No response
Comments:
Used to perform a single point calibration of Temp B.
Temp B sensor must be submerged in LN2. See the Temperature Calibration
Section of this manual.
CALVL
Set Level Offset
Input: CALVL ±XXX.X
Returned:
No response
Comments:
Used to set the liquid level offset.
CALVL?
Level Offset Query
Input: CALVL?
Returned:
±XXX.X
Comments:
Used to query the current level offset value.
CODE?
Global Password Query
Input: CODE?
Returned:
XXXX
Comments:
Used to query the current global password.
HITSTA
Temp A High Alarm Test
Input: HITSTA
Returned:
No response
Comments:
Used to initiate the High Temp A Alarm Test. See the High
Temperature Alarm Test section for more information.