82
32.
Place the sample pads with the sample on
the balance pan. Close the instrument
cover.
33.
Press START.
As the analysis begins, the flashing microwave
indicators in the balance icon appear and the
time begins counting (up for Constant Weight or
down for Set Time). The sample weight
decreases as the moisture is removed.
During the drying time, the operation key below
SOLIDS or MOISTURE may be pressed to
display % solids or % moisture.
Note: Press the operation key below STOP
TEST to end the analysis.
When the weight loss is equal to or less than the
selected parameter, five short beeps will be
heard, and the Data Results screen will appear,
displaying either the % moisture or % solids as
selected.
Note: Press READY to analyze additional
samples using the same method.
Note: Press the operation key below FORM
FEED to advance paper from the internal printer.
34.
Press the operation key below DATA to
display the analysis data.
35.
Press the operation key below SAMPLE
WEIGHTS to view sample weight
information.
XXXXXX
Power:
xxx %
Temp (xxx):
xx.x
Time:
xx:xx
Wt: x.xxxxg
Press sample and pads on balance
and press START.
EXTERNAL
ID
SYRINGE
1 2 3 4 5
T
XXXXX
Sample ID:
xxxxxxxx
Power:
xxx%
Temp (xxx):
xx.x
Time:
xx:xx
Moisture: xx.xx%
Wt: x.xxxxg
STOP
SOLIDS
TEST
XXXXX
XX.XX%M
Press READY to continue.
MAIN
FORM
MENU
FEED
DATA
Data Results
Sample ID:
xxxxxx
Dry Time:
xx:xx
Moisture:
xx.xx%
Solids:
xx.xx%
Press READY to continue.
MAIN
FORM SAMPLE
MENU
FEED
WEIGHTS PRINT