75
27.
Lift the instrument cover. Remove the
syringe from the balance pan.
28.
Press READY.
29.
Place the sample pads with the sample on
the balance pan. Close the instrument
cover.
30.
Press START.
The instrument reads and records the initial
weight of the sample prior to beginning the
analysis.
As the analysis begins, the flashing microwave
indicators in the balance icon appear and the
time begins counting (up for Constant Weight or
down for Set Time). The sample weight
decreases as the moisture is removed.
During the drying time, the operation key below
SOLIDS or MOISTURE may be pressed to
display % solids or % moisture.
Note: Press the operation key below STOP
TEST to end the analysis.
Syringe Weigh (Int)
Remove syringe and press READY.
ABORT
XXXXXX
Power:
xxx %
Temp (xxx):
xx.x
Time:
xx:xx
Wt: x.xxxxg
Place sample and pads on
balance and press START.
EXTERNAL
ID
SYRINGE
1 2 3 4 5
XXXXX
Sample ID:
xxxxxxxx
Power:
xxx%
Temp (xxx):
xx.x
Time:
xx:xx
Moisture: xx.xx%
Wt: x.xxxxg
STOP
SOLIDS
TEST
READING BALANCE . . .
T