105
The instrument reads and records the initial
weight of the sample prior to beginning the
analysis.
As the analysis begins, the flashing microwave
indicators in the balance icon appear and the
time begins counting (up for Constant Weight or
down for Set Time). The sample weight
decreases as the moisture is removed.
During the drying time, the operation key below
SOLIDS or MOISTURE may be pressed to
display % solids or % moisture.
When the weight loss is equal to or less than the
selected parameter, five short beeps will be
heard, and the Data Results screen will appear,
displaying either the % moisture or % solids as
selected.
Note: Press the operation key below FORM
FEED to advance paper from the internal printer.
37.
Press READY to redry the sample or to
analyze additional samples using the same
method.
38.
Lift the cover of the SMART System
5
.
Remove the pads from the balance pan.
39.
Fold the square pads in half and place them
in the extraction chamber of the Fat
Extraction System. Place the round pad
(smooth side up) in the recess at the top of
the extraction chamber. Close and latch the
lid of the extraction chamber.
WARNING
Failure to insert or properly position the
round filter pad in the recess at the top
of the extraction chamber could result
in operator exposure to solvent due to
leakage or possible spraying and/or
damage to the Fat Extraction System,
resulting from solid matter entering the
distillation system.
READING BALANCE . . .
XXXXX
Sample ID:
xxxxxxxx
Power:
xxx%
Temp (xxx):
xx.x
Time:
xx:xx
Ratio:
x.xxxx
Moisture: xx.xx%
Wt: x.xxxxg
STOP
SOLIDS
TEST
XXXXX
XX.XX%M
Press READY to continue.
MAIN
FORM
MENU
FEED
DATA
XXXXXX
Pwer:
xxx %
Temp (xxx): xx.x
Time:
xx:xx
Ratio:
x.xxxx
Wt: x.xxxxg
Placepads on balance and
press TARE.
ID
REDRY
1 2 3 4 5