TARING BALANCE . . .
23.
Lift the top cover of the SMART System
5™. Place two glass fiber sample pads on
the balance pan. Close the top cover.
24.
Press TARE.
25.
Wait for the instrument to zero the weight
of the sample pads. The Quick Test screen
will reappear when the tare function is
complete.
Note: The “T” in the lower left corner of the
balance icon indicates that the weight of the
sample pads has been tared.
26.
Lift the top cover. Remove the sample pads
from the balance pan. Apply the sample in
a thin, even layer to one of the sample
pads. Quickly and gently place the sample
pads back on the balance pan. Close the top
cover.
27.
Press START.
The instrument reads and records the initial
weight of the sample prior to beginning the
analysis.