113
27.
Lift the cover of the SMART System5.
Remove the pads from the balance pan.
28.
Fold the square pads in half and place them
in the extraction chamber of the Fat
Extraction System. Place the round pad
(smooth side up) in the recess at the top of
the extraction chamber. Close and latch the
lid of the extraction chamber.
WARNING
Failure to insert or properly position the
round filter pad in the recess at the top
of the extraction chamber could result
in operator exposure to solvent due to
leakage or possible spraying and/or
damage to the Fat Extraction System,
resulting from solid matter entering the
distillation system.
AVERTISSEMENT
L
’
installation incorrecte du filtre rond
(coussinet) situé sur la partie supérieure
de la chambre d
’
extraction peut exposer
l
’
opérateur à une possible fuite ou vapo-
risataion du solvent. De plus, le systeme
d
’
extraction des matières grasses peut
ètre endommagé résultant du passage
des matières solides dans le système de
distillation.
29.
Press the Run/Stop button on the extractor
to start the extraction cycle. When the
extraction is complete, a beep tone will
sound, and the light in the Run/Stop button
will flash. While the extractor cup is still in
the inverted position, unlatch the lid and
remove the round sample pad on which the
chopped square pads and extracted and
dried sample have been deposited. Latch
the cup lid. The extractor cup will automat-
ically rotate 180 degrees to the upright
position.
30.
Place one round and two square glass fiber
sample pads on the balance pan (round pad
on bottom) of the SMART System
5
. Close
the instrument cover.
31.
Press TARE.
Modified Moisture/Fat
Sample Wt: xx.xxxg
Place sample in Fat Extractor
chamber. Place two new square
pads on balance and press TARE.
ABORT