29
Quick Test
1. POWER: 0%
2. DELTA WEIGHT: 0.0 mg
3. DELTA TIME: 0 secs
4. MAX TIME: 10 mins
5. MAX TEMP: 110 C
6. MIN WT RANGE: 2.00 g
7. MAX WT RANGE: 4.00 g
8. WT COMPENSATION: OFF
Input power, press ENTER.
Entry:
000
SET
TIME
CEM Main Menu
1. QUICK TEST
2. EDIT/CREATE METHOD
3. LOAD METHOD
4. SETUP
5. PRINT
6. STATISTICS
Press item number to select.
METHOD - QUICK TEST
Quick Test
Quick Test permits entry of parameters for a one-
time analysis or method development. The Quick
Test method permits two sample analysis
procedures – Constant Weight and Set Time.
Constant Weight
Constant weight permits the operator to dry a
sample until a constant weight is achieved.
Dryness is specified by defining a maximum
acceptable weight loss over a specified time
interval. During the specified time, when the
weight loss is equal to or less than that which
was specified, the analysis stops and results are
calculated.
1.
With the CEM Main Menu displayed, press
“1” to activate Quick Test.
2.
Press “1” to select and enter power.
3.
Using the numeric keys, enter the power
level (1 - 100%).
4.
Press ENTER.
5.
Press “2” to select and enter delta weight.
Note: To edit any of the parameters, press the
numeric key of the parameter to be edited and
use the numeric keys to enter the new parameter.
Then, press ENTER.
6.
Using the numeric keys, enter the delta
weight (.1 - .9). Normally a weight loss
differential of 0.2 - 0.5 mg is used.
7.
Press ENTER.
Quick Test
1. POWER: 0%
2. DELTA WEIGHT: 0.0 mg
3. DELTA TIME: 0 secs
4. MAX TIME: 10 mins
5. MAX TEMP: 110 C
6. MIN WT RANGE: 2.00 g
7. MAX WT RANGE: 4.00 g
8. WT COMPENSATION: OFF
Press item number to select.
SET
TIME
Quick Test
1. POWER: XXX%
2. DELTA WEIGHT: 0.0 mg
3. DELTA TIME: 0 secs
4. MAX TIME: 10 mins
5. MAX TEMP: 110 C
6. MIN WT RANGE: 2.00 g
7. MAX WT RANGE: 4.00 g
8. WT COMPENSATION: OFF
Input delta weight, press ENTER.
Entry: 0.
0
SET
TIME
➞
➞