79
Note: The method test screen appears indicating
the power, temperature, and time.
16.
If Autotrack is turned off and a method
identification name or number is required,
press the operation key below ID.
17.
Using the operation keys below the arrows,
position the cursor on (select) the first
number or letter of the sample identifi-
cation. Press ENTER. Continue to position
the cursor on each letter or number and
press ENTER until the identification is
complete.
Note: If using an identification number only, use
the numeric keys to enter the number.
18.
Press READY to return to the test screen.
19.
Press the operation key below
EXTERNAL SYRINGE to engage the
external balance and activate the external
balance syringe weigh menu.
Note: If “External Balance” does not appear on
the screen as an option, return to the Main Menu
screen; press “4 - Setup”; press “1 - System
Options”;press “4 - External Balance”; then
press the applicable number to select an external
balance.
20.
Lift the cover of the SMART System
5
.
Place two glass fiber sample pads on the
balance pan. Close the instrument cover.
Press TARE.
XXXXXX
Power:
xxx %
Temp (xxx):
xx.x
Time:
xx:xx
Wt: x.xxxxg
Place pads on balance and
press TARE.
ID
1 2 3 4 5
SAMPLE ID:
0
1 2 3 4 5 6 7 8 9 A B
C D E F G H I J K L M N
O P Q R S T U V W X Y Z
Space
Delete
Highlight choice, press ENTER.
Press READY to continue.
➞
➞
➞
➞
XXXXXX
Power:
xxx %
Temp (xxx):
xx.x
Time:
xx:xx
Wt: x.xxxxg
Place pads on balance and
press TARE.
EXTERNAL
ID
SYRINGE
1 2 3 4 5
Syringe Weigh (Ext)
Place pads on internal
balance and press TARE.
ABORT