52
Vessel Failure Cleanup Procedure
WARNING
Proper precautions must be taken to avoid contact with acids or acid vapors. Protective gear should be
worn as outlined in the user’s safety program for hazardous materials and the reagent manufacturer’s
material safety data sheet. Refer to these guidelines for proper handling and disposal of reagents.
1. Turn the Discover SP-D Clinical off.
2. Disconnect the instrument from the AC power source prior to performing any service procedure.
3. Remove the attenuator assembly by rotating the attenuator counterclockwise.
4. Remove the spill cup.
5. If the spill cup contains glass particles or shards, ensure that the glass is placed into a proper waste container.
6. Rinse the spill cup. Once all debris is rinsed from the spill cup, check for cracks and/or damage. Wipe the spill cup
with an alcohol wipe or equivalent.
Note:
If the spill cup is damaged, replace the spill cup.
7. Check the cavity walls for debris. Remove any debris and wipe down the walls with an alcohol wipe or equivalent.
NOTE:
If the cavity liner is damaged, contact CEM Service at
800-726-5551
or
704-821-7015
or email
.
8. Examine the bottom of the spill cup to ensure that the IR lens cover is in position and undamaged. If the IR lens is
not in position or damaged, call CEM Service. If the IR lens is dirty, use a cotton swab and alcohol to clean the
lens.