40
NOTE:
If the spill cup appears damaged or cracked, contact CEM Corporation at
800-726-5551
or
704-821-7015
or email
to order a replacement part. The replacement part number for the standard spill cup is 162425.
6. Wipe the cavity liner (walls) with an alcohol wipe or equivalent.
7. Inspect the cavity liner for foreign debris or charring.
NOTE:
If charring is observed on the cavity liner, contact CEM Service at
800-726-5551
or
704-821-7015
or email
.
8. Inspect the IR lens for foreign debris and/or damage. If necessary, use a swab and alcohol to clean the IR lens,
using caution not to scratch the IR lens.
NOTE:
If the IR lens is in proper position, it will reflect an image similar to a mirror. If the IR lens appears to be damaged
(ie. scratched), contact CEM Service at
800-726-5551
or
704-821-7015
or email
.
9. If the spill cup is not damaged, install it once it has dried.
NOTE:
Replacement of the spill cup will require calibration of the IR as outlined in this manual.
10. Replace the attenuator and continue operation.
ActiVent
The ActiVent should be cleaned on a weekly basis or when venting occurs. There are two potential cleaning procedures,
as outlined below.
APM Cleaning
1. If applicable, remove the Discover from underneath the Explorer autosampler module.
Note
: Refer to the Explorer Operation Manual for instructions on removal of the Discover module.
2. Press the HOME key until the “Discover Home” screen appears.
3. Use the “+/-” keys to highlight “System Setup.”
4. Press the ENTER key. A new screen will appear.