28
User Profiles
The User Profiles option allows the user to set defaults when creating methods, edit/change the password and edit
system parameters. There are 4 pages within the “User Profiles” page.
1. From the “User Setup” screen, use the “+/-” keys to highlight “User Profiles.”
2. Press the ENTER key.
User Profiles
The “User Profiles” screen is the 1
st
page of the “User Profiles” menu. This page allows the user to clear or change
the password.
Clear Password
1. Use the “+/-” keys to highlight “Clear Password.” The user will be prompted to clear the password - yes or no.
2. Use the ARROW keys to highlight either “Yes” to clear the password or “No” to leave the password as is.
3. Press the ENTER key.
Change Password
1. Use the “+/-” keys to highlight “Change Password.” A new screen will appear.
2. Use the ARROW and “+/-” keys to highlight the first letter or number of the password.
NOTE:
The “Shift” key changes the letters and/or symbols to lowercase.
3. Press the ENTER key to accept the first letter.
4. Continue highlighting the letters or numbers of the password and press the ENTER key for each letter/number until
the password has been entered.
5. Once the password is entered, use the ARROW and “+/-” keys to highlight “OK.”
User Profile Defaults -1
“User Profile Default” is the 2
nd
page of the “User Profile” menu. The “User Profile Defaults” page allows the user to
program a default Hold Time, Ramp Time, Temperature and Microwave Power.
Hold Time
“Hold Time” is the default amount of time defined by the user for the instrument to maintain the control temperature or
pressure. The instrument default hold time is five (5) minutes. When creating a new method, the hold time can be
changed from the default hold time programmed in the “User Profiles.”
1. Use the “+/-” keys to highlight hold time.
2. Press the ENTER key.
3. Use the ARROW keys to toggle between hours, minutes and seconds. Use the “+/-” keys to increase and decrease
the hold time (0 - 99 hours, 59 minutes, 59 seconds).
4. Press the ENTER key to accept the default hold time.