29
Ramp Time
“Ramp Time” is the default amount of time defined by the user for the instrument to reach the temperature or pressure.
The instrument default ramp time is ten (10) minutes. The Discover System will achieve the set temperature as fast as
possible. The ramp time is a maximum amount of time that power will be applied before switching to the hold time. When
creating a new method, the ramp time can be changed from the default ramp time programmed in the “User Profiles.”
1. Use the “+/-” keys to highlight “Ramp Time.”
2. Press the ENTER key.
3. Use the ARROW keys to toggle between hours, minutes and seconds. Use the “+/-” keys to increase and decrease
the hold time (0 - 99 hours, 59 minutes, 59 seconds).
4. Press the ENTER key to accept the default ramp time.
Temperature
“Temperature” is the user defined control temperature. The default temperature is 125°C. When creating a new method,
the temperature can be changed from the default temperature programmed in the “User Profiles.”
1. Use the “+/-” keys to highlight “Temperature.”
2. Press the ENTER key.
3. Use the “+/-” keys to increase and decrease the temperature from 0-220°C.
4. Press the ENTER key to accept the default temperature.
Power
Microwave power is the user defined maximum power used during a method. The default microwave power is 150 watts.
When creating a new method the microwave power can be changed from the default power programmed in the “User
Profiles.”
1. Use the “+/-” keys to highlight “Temperature.”
2. Press the ENTER key.
3. Use the “+/-” keys to increase and decrease the microwave power from 0-300 watts.
4. Press the ENTER key to accept the default microwave power.
User Profile Defaults- 2
From the 1
st
“User Profile Default” screen, press the right arrow key one time to access the 2
nd
“User Profile Default”
screen. This is also the 3
rd
page of the “User Profile” menu. This page allows the user to program default Stirring, Pre-Mix
Time, Pressure and Pressure Units.
Stirring
“Stirring” is the user defined stirring speed for the method. The default stirring speed is high (this is also the recommended
speed when performing reactions in the 10 mL vials). When creating a new method the stirring speed can be changed
from the default stirring programmed in the “User Profiles.”
1. Use the “+/-” keys to highlight “Stirring.”
2. Press the ENTER key.
3. Use the “+/-” keys to increase and decrease the stirring between “High,” “Med,” “Low” and “Off”.
4. Press the ENTER key to accept the default stirring speed.