![Cellink BIO X6 User Manual Download Page 27](http://html1.mh-extra.com/html/cellink/bio-x6/bio-x6_user-manual_2580414027.webp)
BIO X6 User Manual v. 1.0
27
3.5 Calibration trimming
The BIO X6 has a protocol for running an auto calibration trimming procedure. This
procedure consists of an automatic calibration followed by a manual calibration of all six
printheads. The whole process takes about 15 minutes.
Auto calibration trimming must be performed before using the system for the first time, to
ensure that the automatic calibration is correctly aligned.
1. Tap
> Printing utilities > Machine Calibration > Perform auto calibration trimming
2. Attach all 6 pneumatic printheads. Use the same kind of nozzle/needle for all
printheads.
3. Attach the surface probe and the metal reflectors.
Objects left on the printbed may damage the printer. Ensure the printbed is
empty without Petri dish, well plate, glass slice, or debris.
4. Tap
Start calibration
to start the automatic calibration.
5. Tap
Continue
to proceed to the manual calibration of each printhead.
6. Follow the step-by-step instructions shown on the screen.
Figure 7 Auto calibration trimming
Figure 6 Auto calibration trimming, continued
Summary of Contents for BIO X6
Page 1: ...BIO X6 User Manual v 1 0 1 BIO X6 User Manual...
Page 8: ...BIO X6 User Manual v 1 0 8 01 Safety...
Page 14: ...BIO X6 User Manual v 1 0 14 02 Specifications...
Page 19: ...BIO X6 User Manual v 1 0 19 03 Getting started...
Page 29: ...BIO X6 User Manual v 1 0 29 04 Bioprinting...
Page 48: ...BIO X6 User Manual v 1 0 48 Figure 22 Custom well plate option...
Page 68: ...BIO X6 User Manual v 1 0 68 Figure 42 Manual calibration glass slide...
Page 94: ...BIO X6 User Manual v 1 0 94 05 Gcode Editor...
Page 104: ...BIO X6 User Manual v 1 0 104 Figure 77 Error message Command not supported...
Page 105: ...BIO X6 User Manual v 1 0 105 06 Draw Print...
Page 117: ...BIO X6 User Manual v 1 0 117 07 Shape Generator...
Page 122: ...BIO X6 User Manual v 1 0 122 08 Print report...
Page 126: ...BIO X6 User Manual v 1 0 126 Figure 97 Protocol setup continued...
Page 130: ...BIO X6 User Manual v 1 0 130 09 Coaxial printing...
Page 134: ...BIO X6 User Manual v 1 0 134 10 Maintenance...
Page 146: ...BIO X6 User Manual v 1 0 146 11 Terms and conditions...
Page 166: ...BIO X6 User Manual v 1 0 166 12 Printhead manuals...
Page 175: ...BIO X6 User Manual v 1 0 175...
Page 176: ...BIO X6 User Manual v 1 0 176 Create the future of medicine...