10.3
Manually cleaning or sterilizing the BIO X6
Always turn off and unplug the BIO X6 before cleaning it with ethanol. Wear
gloves to avoid direct contact with the bioprinter.
The BIO X6 has a chemical-resistant coating. Clean or sterilize it with >70%
(v/v) ethanol. Spray the surface and wipe it with a soft cloth.
After using ethanol, wait at least 60 seconds before starting the BIO X6. Ethanol
is flammable, and this ensures that all remaining ethanol vapor has been
ventilated. Failure to wait can cause equipment damage and personal injury.
Task
Action
Regularity
Check
Dates
Clean door
(in/out)
Clean with a smooth cloth. Do not use
harsh solvents. The glass is sensitive to
acetone and other chemicals.
Every week
Clean top plate
Clean carefully, avoid pushing
dust/particles down to the pre-filter.
Every week
Clean inside
chamber
Remove any dirt on the rail or the lead
screw with a clean, lint-free cloth or with
your finger (with gloves on). After 6
months, it can be relubricated with clear
fluoropolymer grease.
Every week
Clean printbox
Clean with a smooth cloth, avoid using
hash solvents.
Every time it
is used
Clean printbed
(top plate)
Clean the surface with a cloth and
ethanol
>70% (v/v).
Every time it
is used
Clean printbed
(lower white part)
Clean with a smooth cloth, avoid using
harsh solvents.
Every week
Summary of Contents for BIO X6
Page 1: ...BIO X6 User Manual v 1 0 1 BIO X6 User Manual...
Page 8: ...BIO X6 User Manual v 1 0 8 01 Safety...
Page 14: ...BIO X6 User Manual v 1 0 14 02 Specifications...
Page 19: ...BIO X6 User Manual v 1 0 19 03 Getting started...
Page 29: ...BIO X6 User Manual v 1 0 29 04 Bioprinting...
Page 48: ...BIO X6 User Manual v 1 0 48 Figure 22 Custom well plate option...
Page 68: ...BIO X6 User Manual v 1 0 68 Figure 42 Manual calibration glass slide...
Page 94: ...BIO X6 User Manual v 1 0 94 05 Gcode Editor...
Page 104: ...BIO X6 User Manual v 1 0 104 Figure 77 Error message Command not supported...
Page 105: ...BIO X6 User Manual v 1 0 105 06 Draw Print...
Page 117: ...BIO X6 User Manual v 1 0 117 07 Shape Generator...
Page 122: ...BIO X6 User Manual v 1 0 122 08 Print report...
Page 126: ...BIO X6 User Manual v 1 0 126 Figure 97 Protocol setup continued...
Page 130: ...BIO X6 User Manual v 1 0 130 09 Coaxial printing...
Page 134: ...BIO X6 User Manual v 1 0 134 10 Maintenance...
Page 146: ...BIO X6 User Manual v 1 0 146 11 Terms and conditions...
Page 166: ...BIO X6 User Manual v 1 0 166 12 Printhead manuals...
Page 175: ...BIO X6 User Manual v 1 0 175...
Page 176: ...BIO X6 User Manual v 1 0 176 Create the future of medicine...