LSS-2404
25 |
P a g e
Recipe Settings
Trigger Type (16)
Triggered One-shot (default
)
–
Executes one triggered recipe per rising edge on the Trigger In signal, regardless
of the signal’s length. The signal must return to the low state before the LSS
can be triggered again.
Triggered Continuous
–
Executes a triggered recipe on the Trigger In signal rising edge and continues repeating
the recipe as long as the signal level remains in the high state. The execution will end at the completion of the
last recipe where the trigger fell to the low state.
Free-run
–
Recipe repeats continuously without regard to the Trigger. They start and stop when the Free-run
mode is enabled and disabled.
Trigger Event (17)
Sets the trigger to initiate a recipe, sequence or frame. These can only be selected if the
‘Trigger Type’ is set to
One-Shot.
Series
–
Each trigger will run through the entire programmed recipe until completion.
Sequence
–
Each trigger will run through one sequence. If the sequence has multiple executions all executions
will be completed from one trigger. If a recipe has multiple sequences the next trigger will trigger the next
sequence. Once all the sequences have run the next trigger will trigger the first sequence again.
Frame
–
Each trigger will initiate one frame at a time. If there is one sequence, the triggers will continue to cycle
through the frames of the sequence. If there are multiple sequences, once all the frames in a sequence have
been triggered the next trigger will trigger frame one in the next sequence.
Ground Trigger (18)
If checked, trigger level voltage is ignored. Use this for contact closure type trigger.
Trigger Level (15)
Sets the voltage level of the Trigger In signal at which the LSS will trigger. The default setting (9.6 V) will
generally work well for 12-24 V trigger signals. Set this value lower to work with 5 V or 3.3 V logic levels. Set the
value lower if the trigger signal is having problems consistently triggering the unit. Set the value higher to reject
noise or if the unit false triggers.
Trigger Delay (14)
Delays sending the initial output signals for a set time once the input trigger is received. Only delays the first
trigger in the recipe
so it can only be set when ‘Trigger Type’ is ‘One
-
shot and ‘Trigger Event’ is ‘Series’.
Maximum delay is 16,000 milliseconds or microseconds (depending on unit Resolution setting).
Camera Trigger Voltage (12)
Sets the voltage sent to the camera when it triggers
Camera and Trigger Polarity (13/19)
Inverts the logic of the trigger in the software. When set to positive it triggers on rising edge, when negative it
triggers on the falling edge.
Note: the hardwire of the trigger cannot change