USER MAINTENANCE
Optional: Obtain a temperature reference device
. A calibrated digital thermometer with a vacuum-
rated thermocouple feedthrough. The device must be accurate to at least 0.1°C.
Before Starting
1.
The unit must be connected to a power source that meets the requirements in the Installation chapter
(page 25) and turned on.
2.
Optional: secure the reference temperature device sensor probe at the center of the bottom shelf, with
the probe head in direct contact with the shelf surface.
3.
The oven chamber must be sealed and be under vacuum. See the
Place the Chamber Under Vacuum
entry on page 42.
4.
The unit must have adequate time to come up to temperature and stabilize.
Failure to wait will result
in an inaccurate diagnosis
.
•
Allow 120 minutes for the unit to achieve 150°C or 175 minutes to achieve 220°C. The
unit cannot come up to temperature and stabilize faster than these rates.
•
Start the Diagnostic Data Procedure
when the allotted time has passed
, even if the
unit fails to achieve the set point temperature.
Summary of Contents for CVO-2-HT
Page 1: ...Vacuum Ovens CVO 2 HT CVO 5 HT Installation Operation Manual...
Page 6: ...6 P a g e cascadesciences com Tel 503 847 9047 TABLE OF CONTENTS...
Page 8: ...8 P a g e cascadesciences com Tel 503 847 9047 CERTIFICATIONS...
Page 12: ...12 P a g e cascadesciences com Tel 503 847 9047 UNIT SPECIFICATIONS...
Page 16: ...16 P a g e cascadesciences com Tel 503 847 9047 INTRODUCTION...
Page 22: ...22 P a g e cascadesciences com Tel 503 847 9047 RECEIVING YOUR OVEN...
Page 32: ...32 P a g e cascadesciences com Tel 503 847 9047 GRAPHIC SYMBOLS...
Page 71: ...71 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS...
Page 72: ......