24 |
P a g e
cascadesciences.com
|
Tel. 503 847-9047
INSTALLATION
R
EQUIRED
A
MBIENT
C
ONDITIONS
When selecting a location to install the unit, consider all environmental conditions that can adversely
impact its temperature performance. These include:
•
Proximity to other ovens, autoclaves, and any device that produces significant radiant heat
•
Heating and cooling vents or other sources of fast-moving air currents
•
High-traffic areas
•
Direct sunlight
This oven is built for use indoors at room temperatures between
15
°
C and 40
°
C (59
°
F and 104
°
F)
, at no
greater than
80% Relative Humidity
(at 25
°
C / 77
°
F). Operating outside these conditions may adversely
affect the oven temperature performance.
R
EQUIRED
C
LEARANCES
These clearances are required to provide air flows for ventilation and cooling.
6 inches (152 mm)
of clearance is required on the sides.
12 inches (305 mm)
of headspace clearance is required between the top of the unit and any
overhead partitions.
Do not place objects on top of the oven.
A KF-25 vacuum port is located on the back of the oven. Leave sufficient clearance for end-users to
safely access this port.
6” (152 mm)
Fan
KF-25 Port
6” (152 mm)
Door Swing
12” (305 mm)
12” (305 mm)
12” (305 mm)
6” (152 mm)
CVO-2-HT: 20.5” (521 mm)
CVO-5-HT: 27.0” (686 mm)
Summary of Contents for CVO-2-HT
Page 1: ...Vacuum Ovens CVO 2 HT CVO 5 HT Installation Operation Manual...
Page 6: ...6 P a g e cascadesciences com Tel 503 847 9047 TABLE OF CONTENTS...
Page 8: ...8 P a g e cascadesciences com Tel 503 847 9047 CERTIFICATIONS...
Page 12: ...12 P a g e cascadesciences com Tel 503 847 9047 UNIT SPECIFICATIONS...
Page 16: ...16 P a g e cascadesciences com Tel 503 847 9047 INTRODUCTION...
Page 22: ...22 P a g e cascadesciences com Tel 503 847 9047 RECEIVING YOUR OVEN...
Page 32: ...32 P a g e cascadesciences com Tel 503 847 9047 GRAPHIC SYMBOLS...
Page 71: ...71 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS...
Page 72: ......