USER MAINTENANCE
Diagnostic Data Procedure
P
ump On and Running?
Yes or no?
V
ent Valve Closed?
The vent (backfill inlet port) must be closed before applying vacuum to the chamber. Failure to do so
may result in damage to your vacuum pump.
V
acuum Valve Open?
The vacuum valve must be open to allow a connected vacuum supply to evacuate the oven
chamber.
D
isplay Reading?
Record the chamber pressure level showing on the Vacuum Gauge display.
Reminder
: Make sure to record the correct unit of measurement for the chamber pressure. See
the Vacuum Gauge Operations topic on page 46 to verify the current unit of measurement.
Vacuum Valve
AND
Inches of Mercury