OPERATION
V
ACUUM
G
AUGE
O
PERATIONS
Change the Unit of Measurement
1.
Place the vacuum gauge in its adjustment mode.
a.
Press and hold the “M” button for approximately 3 seconds
•
The display will begin to blink and show a unit of measurement
2.
Use the arrow keys to scroll between units.
3.
Exit the adjustment mode.
a.
Press and hold the “M” button for approximately 3 seconds
•
The display will cease blinking and show the current chamber
pressure.
Units of Measurement – Display Characters
kPa
Kgf/cm
2
bar
psi
mmHg
inHg
mmH
2
O
Zeroing the Gauge
As set at the factory, the vacuum gauge shows a reading of 0 inches of mercury (inHg) when the
chamber is at ambient (room) pressure. The display was set at near sea level.
If the gauge does not show 0 inHg when the chamber is at room atmospheric pressure, perform the
following steps.
1.
With the chamber door open
, press and hold both the
Up
and
Down
arrow buttons.
2.
Release the buttons when the display shows 0.0.
See page 11 for the zero equivalent for units of measurement other than inHg.
Inches of Mercury
Kilopascals
Both
Summary of Contents for CVO-2-HT
Page 1: ...Vacuum Ovens CVO 2 HT CVO 5 HT Installation Operation Manual...
Page 6: ...6 P a g e cascadesciences com Tel 503 847 9047 TABLE OF CONTENTS...
Page 8: ...8 P a g e cascadesciences com Tel 503 847 9047 CERTIFICATIONS...
Page 12: ...12 P a g e cascadesciences com Tel 503 847 9047 UNIT SPECIFICATIONS...
Page 16: ...16 P a g e cascadesciences com Tel 503 847 9047 INTRODUCTION...
Page 22: ...22 P a g e cascadesciences com Tel 503 847 9047 RECEIVING YOUR OVEN...
Page 32: ...32 P a g e cascadesciences com Tel 503 847 9047 GRAPHIC SYMBOLS...
Page 71: ...71 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS...
Page 72: ......