USER MAINTENANCE
Diagnostic Data Procedure – SDRAP Questions
Record the answers in the log on page 62.
S
et Point?
What is the current temperature set point
?
D
isplay?
What chamber temperature is presently showing on the temperature display?
R
eference?
Optional: What temperature is the reference device presently showing for the chamber
temperature?
A
mbient?
What is the room temperature?
For best results, measure the temperature in the same section of
the room where the unit is located. Do not place your thermometer on the unit.
P
ilot Lights?
1) Is the heating active indicator on the control panel flashing or otherwise illuminating, Y/N?
2) Is the High Limit cutoff active or has it activated recently, Y/N?
Chamber Temperature in Red
Set Point in Green
“2” indicates the
controller is calling for
power to the element
Alternating alert screens flash when the high limit heating cutoff is active.
Summary of Contents for CVO-2-HT
Page 1: ...Vacuum Ovens CVO 2 HT CVO 5 HT Installation Operation Manual...
Page 6: ...6 P a g e cascadesciences com Tel 503 847 9047 TABLE OF CONTENTS...
Page 8: ...8 P a g e cascadesciences com Tel 503 847 9047 CERTIFICATIONS...
Page 12: ...12 P a g e cascadesciences com Tel 503 847 9047 UNIT SPECIFICATIONS...
Page 16: ...16 P a g e cascadesciences com Tel 503 847 9047 INTRODUCTION...
Page 22: ...22 P a g e cascadesciences com Tel 503 847 9047 RECEIVING YOUR OVEN...
Page 32: ...32 P a g e cascadesciences com Tel 503 847 9047 GRAPHIC SYMBOLS...
Page 71: ...71 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS...
Page 72: ......