USER MAINTENANCE
End Calibration procedure
Calibration continued
8
Allow the oven to stabilize after achieving the temperature
set point using the offset display value.
9
Once the chamber has stabilized, compare the reference
temperature device and oven temperature display readings.
•
If the readings are the same or the difference
between the two falls within the acceptable range
of your protocol, the display is accurately showing
the chamber temperature.
The calibration
procedure is now complete
.
-OR-
a.
See step 10 if the difference falls outside the
acceptable range of your protocol again.
Reference Device
10
Repeat steps 2 – 9 up to two more times.
•
You may skip Step 3 by leaving the controller
unlocked until the unit is successfully calibrated.
•
Three calibration attempts may be required to
successfully calibrate ovens that are more than ± 2°C
out of calibration.
Note:
Always relock the temperature controller after a successful
calibration has been carried out. This safeguards against a user
accidentally changing the controller configuration file and
interfering with the functionality of the unit.
Reference Device
If the temperature reading difference still falls outside your protocol after 3 calibration attempts, contact
your distributor or
Technical Support
for assistance.
Summary of Contents for CVO-2-HT
Page 1: ...Vacuum Ovens CVO 2 HT CVO 5 HT Installation Operation Manual...
Page 6: ...6 P a g e cascadesciences com Tel 503 847 9047 TABLE OF CONTENTS...
Page 8: ...8 P a g e cascadesciences com Tel 503 847 9047 CERTIFICATIONS...
Page 12: ...12 P a g e cascadesciences com Tel 503 847 9047 UNIT SPECIFICATIONS...
Page 16: ...16 P a g e cascadesciences com Tel 503 847 9047 INTRODUCTION...
Page 22: ...22 P a g e cascadesciences com Tel 503 847 9047 RECEIVING YOUR OVEN...
Page 32: ...32 P a g e cascadesciences com Tel 503 847 9047 GRAPHIC SYMBOLS...
Page 71: ...71 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS...
Page 72: ......