USER MAINTENANCE
U
NLOCKING THE
T
EMPERATURE
C
ONTROLLER
The controller must be unlocked to access the Operations menu and enter calibration offsets.
The oven temperature controller is software locked at the factory to ensure the integrity of its
configuration file. This safeguards against end-users accidentally altering the oven functionality or
safe operating bounds.
Backing Up the Configuration File
The manufacturer recommends saving the controller configuration file prior to making any changes
to Operations options. See the Configurator software description in the Data Port description of the
user manual. This will allow you to restore the configuration file in the event a change is made that
adversely affects the operation of the oven.
Continued next page
1.
Starting on the Home page, navigate to the Lock menu
Simultaneously press and hold
both
the
Reset
and
Advance
buttons for approximately 8 – 9
seconds.
Note:
If the top red line shows the “CUSt”
Custom option, use the
Up
or
Down
arrow keys
to scroll to the “Loc” Security Setting option.
Then push the Advance button as per Step 2.
2.
Advance to the lock “LoC.o” parameter
3.
Adjust the LoC.o setting from 3 to 2
Press and
Hold Both
Push Once
Summary of Contents for CVO-2-HT
Page 1: ...Vacuum Ovens CVO 2 HT CVO 5 HT Installation Operation Manual...
Page 6: ...6 P a g e cascadesciences com Tel 503 847 9047 TABLE OF CONTENTS...
Page 8: ...8 P a g e cascadesciences com Tel 503 847 9047 CERTIFICATIONS...
Page 12: ...12 P a g e cascadesciences com Tel 503 847 9047 UNIT SPECIFICATIONS...
Page 16: ...16 P a g e cascadesciences com Tel 503 847 9047 INTRODUCTION...
Page 22: ...22 P a g e cascadesciences com Tel 503 847 9047 RECEIVING YOUR OVEN...
Page 32: ...32 P a g e cascadesciences com Tel 503 847 9047 GRAPHIC SYMBOLS...
Page 71: ...71 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS...
Page 72: ......