25
|
P a g e
cascadesciences.com
|
Tel. 503 847-9047
INSTALLATION
P
OWER
S
OURCE
R
EQUIREMENTS
When selecting a location for the unit, verify each of the following requirements is satisfied.
Power Source
: The wall power outlet must meet the power requirements listed on the unit data
plate.
Model
AC Voltage
Amperage
Frequency
CVO-2-HT
110 – 120
10
50/60 Hz
CVO-5-HT
110 – 120
13
50/60 Hz
•
Wall power sources must be protective earth grounded and
single phase
.
•
Wall power sources must conform to all national and local electrical codes.
•
Supplied voltage must not vary more than 10% from the data plate rating
. Damage to the unit
may result if the supplied voltage varies more than 10%.
•
The recommended wall circuit breakers for these units are 15 amps.
•
Use a separate circuit to prevent loss of product due to overloading or circuit failure. The circuit
must match or exceed the amperage requirement listed on the unit data plate.
Power Cord
: The unit must be positioned so that all end-users can quickly unplug the oven in the
event of an emergency.
•
The unit comes provided with a 125 volt, 15 amp, 9ft 5 in (2.86m) NEMA 5-15P power cord.
Fuses:
These units each ship with a 16 amp 250V 5x20mm fuse installed in a fuse holder
immediately adjacent to the power cord inlet.
•
The fuse must be installed and intact for the unit to operate.
•
Always find and fix the cause of a blown fuse prior to putting the unit back into operation.
Standard 15-
amp NEMA
5-15 power
outlet.
Summary of Contents for CVO-2-HT
Page 1: ...Vacuum Ovens CVO 2 HT CVO 5 HT Installation Operation Manual...
Page 6: ...6 P a g e cascadesciences com Tel 503 847 9047 TABLE OF CONTENTS...
Page 8: ...8 P a g e cascadesciences com Tel 503 847 9047 CERTIFICATIONS...
Page 12: ...12 P a g e cascadesciences com Tel 503 847 9047 UNIT SPECIFICATIONS...
Page 16: ...16 P a g e cascadesciences com Tel 503 847 9047 INTRODUCTION...
Page 22: ...22 P a g e cascadesciences com Tel 503 847 9047 RECEIVING YOUR OVEN...
Page 32: ...32 P a g e cascadesciences com Tel 503 847 9047 GRAPHIC SYMBOLS...
Page 71: ...71 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS...
Page 72: ......