CareFusion Alaris SE 7130 Directions For Use Manual Download Page 9

1000DF00480 Issue 2                 8/48               

Operating 

Pressure

• 

The pumping pressure alarm system is not designed to provide protection against, or detection  

of, extravasation or tissuing complications which can occur. 

• 

The pump is designed to stop fluid flow under alarm conditions.  Periodic patient monitoring must be 

performed to ensure the infusion is proceeding as expected.  It is a positive displacement delivery system, 

capable of developing positive fluid pressures to overcome widely varying resistances to flow encountered 

in practice, including resistances to flow imposed by small gauge catheters, filters and intra-arterial infusion.  

It is neither designed nor intended to detect infiltrations and will not alarm under infiltration conditions.

Alarm Conditions

• 

Several alarm conditions detected by this pump will stop the infusion and generate visual and 

audible alarms. Users must perform regular checks to ensure that the infusion is progressing 

correctly and no alarms are operating.

Electromagnetic Compatibility and Interference

M

• 

This pump is protected against the effects of external interference, including high energy radio 

frequency emissions, magnetic fields and electrostatic discharge (for example, as generated by 

electrosurgical and cauterising equipment, large motors, portable radios, cellular telephones 

etc.) and is designed to remain safe when unreasonable levels of interference are encountered.

• 

In some circumstances the pump may be affected by an electrostatic discharge through air 

at levels close to or above 15kv; or by radio frequency radiation close to or above 10v/m. If 

the pump is affected by this external interference the pump will remain in a safe mode; the 

pump will duly stop the infusion and alert the user by generating a combination of visual and 

audible alarms. Should any encountered alarm condition persist even after user intervention, it 

is recommended to replace that particular pump and quarantine the pump for the attention of 

appropriately trained technical personnel.

• 

This pump is a CISPR 11 Group 1 Class A device when the model 180 (Flow Sensor) is used and a 

CISPR 11 Group 1 Class B device without the use of the model 180 (Flow Sensor). The pump uses 

RF energy only for its internal function in the normal product offering. In a domestic environment, 

this system may cause radio interference.  Reorienting, relocating or shielding the system, or 

filtering the connection to the public mains network, are examples of steps that can be taken 

to reduce or eliminate interference. However, this pump emits a certain level of electromagnetic 

radiation which is within the levels specified by IEC/EN60601-2-24 and IEC/EN60601-1-2. If the 

pump interacts with other equipment, measures should be taken to minimise the effects, for 

instance by repositioning or relocation.

• 

Therapeutic Radiation Equipment: Do not use the pump in the vicinity of any Therapeutic Radiation 

Equipment. Levels of radiation generated by the radiation therapy equipment such as Linear Accelerator, 

may severely affect functioning of the pump. Please consult manufacturer’s recommendations for 

safe distance and other precautionary requirements. For further information, please contact your local 

CareFusion representative.

• 

Magnetic Resonance Imaging (MRI): The pump contains ferromagnetic materials which are susceptible 

to interference with magnetic field generated by the MRI devices. Therefore, the pump is not considered 

an MRI compatible pump as such. If use of the pump within an MRI environment is unavoidable, then 

CareFusion highly recommends securing the pump at a safe distance from the magnetic field outside the 

identified ‘Controlled Access Area’ in order to evade any magnetic interference to the pump; or MRI image 

distortion. This safe distance should be established in accordance with the manufacturers’ recommendations 

regarding electromagnetic interference (EMI). For further information, please refer to the product technical 

service manual (TSM). Alternatively, contact your local CareFusion representative for further guidance.

• 

Accessories:

 

Do not use any non-recommended accessory with the pump. The pump is tested and compliant 

with the relevant EMC claims only with the recommended accessories. Use of any accessory, transducer or 

cable other than those specified by CareFusion may result in increased emissions or decreased pump 

immunity.

Earth Conductor

d

• 

The pump is a Class I device, therefore must be earthed when connected to an AC power 

supply.

• 

This pump also has an internal power source.

• 

When connected to an external power source, a three-wire (Live, Neutral, Earth) supply must 

be used. If the integrity of the external protective conductor on the AC power cable has been 

compromised, the pump should be disconnected from the AC power source and operated 

utilising the internal battery.

Operating Precautions (continued)

Summary of Contents for Alaris SE 7130

Page 1: ...s Alaris SE Pump Models 7130 7131 and 7230 7231 with Guardrails Safety Software Directions For Use en...

Page 2: ...Multi Dose 18 Multi Step 19 Monitoring Options 20 SmartSite Needle Free System Instructions 24 Alarms Alerts and Prompts 25 Alarms 25 Alerts 26 Prompts 27 Guardrails Event Counter 29 Flow Sensor Oper...

Page 3: ...future updates and reference Information about a Guardrails alert that occurs during use is stored within the Pump and can be accessed using the CQI Reporter The Pump may be operated with or without t...

Page 4: ...SEC KVO OPT 2 3 4 5 6 7 8 9 0 1 POWER RUN HOLD A B A B ml hr ml hr A B A B Latch Pumping Mechanism Pressure Transducer Air in line Detector Air in line Arm Panel Lock Keypad AC power indicator Rate Di...

Page 5: ...key Silences audible alarm or alert for 2 minutes message remains on screen New alarm or alert reinstates the audible tone PRI Primary key Selects Primary mode Channel must be selected if applicable S...

Page 6: ...ts of Council Directive 93 42 EEC as amended by 2007 47 EC T Date of Manufacture t Manufacturer B Not for Municipal Waste e Functional Earth RS 232 RS232 Nurse Call Connector Canadian and U S Certific...

Page 7: ...ing when the pump is turned off Panel lock icon Indicates that panel lock is on Audio Volume Indicator Indicates audio volume for alarms and alerts Low Medium High MNTR Computer Mode Indicator Display...

Page 8: ...fusion set if using glass bottles or semi rigid containers to reduce the partial vacuum formed as the fluid is infused from the container This action will ensure the pump can maintain volumetric accur...

Page 9: ...adio interference Reorienting relocating or shielding the system or filtering the connection to the public mains network are examples of steps that can be taken to reduce or eliminate interference How...

Page 10: ...ibility and pump performance as part of the overall infusion Potential hazards include drug interactions inaccurate delivery rates and pressure alarms and nuisance alarms When delivering an infusion w...

Page 11: ...use The pump will automatically operate from its internal battery if the pump is switched on without being connected to the power supply Should the pump fail to perform correctly replace in its origi...

Page 12: ...the manufacturer s directions for use Use only sets dedicated for use with the Alaris SE Pump 1 Slide Flow Regulator thumb clamp down until an audible click verifies it is in fully closed position 2 S...

Page 13: ...ow Regulator A After set installation verify no fluid is flowing through the infusion set s drip chamber to avoid free flow Before operating pump verify that infusion set is free from kinks and instal...

Page 14: ...ress RUN HOLD key to start infusion 5 To briefly view current profile from RUN HOLD page or during a running infusion press 4 soft key New Guardrails Drug 1 Select a drug name by pressing the adjacent...

Page 15: ...oft key 3 Press soft key next to parameter to be edited Current value is highlighted 4 To enter a new value use numeric keypad then press ENTER key Getting Started continued A Rate LED flashes while i...

Page 16: ...guration settings for the selected profile and the dosing limits for the selected drug to be viewed 1 Press OPTIONS key Options page appears 2 To view additional settings press page soft key Second op...

Page 17: ...nfusion set 6 Press the SEC key Secondary Setup page displays 7 To enter desired Rate use numeric keypad then press ENTER key 8 To enter desired VTBI use numeric keypad then press ENTER key 9 Verify t...

Page 18: ...y or to enter a new vtbi use numeric keypad and press ENTER key 6 To start loading dose infusion press RUN HOLD key 7 To briefly view current profile press 4 soft key A This mode is useful for deliver...

Page 19: ...Alert Option is enabled DOSE COMPLETE ALERT OPTION page appears To select On or Off use soft keys 11 To continue programming press ok soft key 12 To start first dose immediately press ok soft key The...

Page 20: ...e channel on hold press RUN HOLD key 2 To return to review page s press setup soft key A tick mark y next to a step on review page s indicates it has not started Only steps having a y can be edited A...

Page 21: ...stable Pressure designed to monitor IV infusion set site pressure and provide user adjustable pressure alarm limits Used for Precision Flow mode or for high resistance systems such as infusion through...

Page 22: ...pressure limit is exceeded This limit may be set by qualified service personnel from 1 mmHg to 600 mmHg Pressure Limit Maximum Resistance Alert The Resistance Alert provides an early warning of increa...

Page 23: ...timal baseline is maintained upon subsequent activations of RUN HOLD as follows If current system pressure is same or higher than original baseline pressure baseline does not change If current system...

Page 24: ...alibrate based on current system pressure 1 To place channel on hold press channel s RUN HOLD key All infusions connected to channel being baselined must be on hold 2 Press options key 3 Press Set Pre...

Page 25: ...S Use Aseptic Technique 1 Prior to every access swab top of Needle Free Valve port with 70 Isopropyl alcohol 1 2 seconds and allow to dry approximately 30 seconds NOTE Dry time is dependent on tempera...

Page 26: ...ual channel pump only Channel malfunction Turn channel off and then on If problem persists do not use channel Contact qualified service personnel computer link failure RS 232 connection to computer wa...

Page 27: ...nked tubing closed clamp Verify that blue thumb clamp on Flow Regulator has moved to open up position If not reload set Press run soft key to restart infusion primary flow detected during secondary Pu...

Page 28: ...ode 600 ml h in secondary mode Verify and re enter settings New Baseline Set A new Manual Pressure Baseline was successfully set Manual Pressure Baseline feature is on Baseline remains set until a new...

Page 29: ...TBI or lower Loading Dose VTBI as appropriate Stop Timer to Change An invalid key was pressed while timer was running in Multi Dose program Wait several seconds for popup to finish Press stop timer so...

Page 30: ...Events maximum rate maximum weight soft dose limit high or low hard dose limit high or low drug selection cancelled Overrides Total number of Guardrails Soft Limits that are overridden Counter increme...

Page 31: ...in a No Upstream Flow Detected alarm This is because the small rapidly falling drops form a continuous stream which does not trigger the flow sensor In this event unplug the flow sensor from the pump...

Page 32: ...arts If allowable number of restarts is exceeded or when resistance or pressure conditions indicate an occlusion an occlusion downstream alarm occurs Monitoring Options Dynamic monitoring provides cli...

Page 33: ...Library Bolus Dose feature may be used Multi Dose Allows 1 24 doses to be programmed at equally spaced intervals on same pump over a 24 hour period This mode allows delivery of multiple equal doses fr...

Page 34: ...lus Off On Off Dose Rate Calculator Off On Off Loading Dose On On Off Multi Dose Off On Off Multi Dose Alert Off On Off Multi Step Off On Off Panel Lock On On Off Rate Maximum 999 9 ml h 0 1 999 9 ml...

Page 35: ...oint failure does not exceed 1 0 ml at 999 9 ml h Dimensions Nominal 7131 7231 Depth 127mm 127mm Height 218mm 218mm Width 193mm 267mm Weight 3 0 kg 3 8 kg Without pole clamp Without power cord Downstr...

Page 36: ...listed below Rate Accuracy Test Conditions Infusion rate range 0 1 to 999 9 ml h Head height 24 1 in 61 2 5 cm Test solution distilled water Environment temperature 68 8 F 20 4 C Back pressure 0 psi N...

Page 37: ...accuracy is reached after several complete charge discharge recharge cycles in the refresh process It is recommended that the battery be fully charged discharged recharged using the refresh cycle bef...

Page 38: ...y running warm water over it while actuating slider and then thoroughly dried After cleaning the sensor should be allowed to dry fully prior to use Disposal Information on Disposal for Users of Waste...

Page 39: ...ith the cable jacket Connecting to a Computer 1 Press options key 2 To advance to next page press page soft key 3 Press Computer Link soft key 4 Press Monitor soft key 5 Press ok soft key 6 Connect an...

Page 40: ...also be considered The start up curves represent continuous flow rate versus operating time for two hours from the start of the infusion They exhibit the delay in onset of delivery due to mechanical c...

Page 41: ...at 1 mL h initial 0 8 1 1 4 9 4 9 6 2 2 0 12 1 3 9 21 2 12 2 3 2 15 10 5 0 5 10 15 0 5 10 15 20 25 30 Observation Interval min Flow Rate Error Pressure Mode Trumpet Curve at 0 1 mL h 48 hr 21 8 11 9...

Page 42: ...Pressure Mode Trumpet Curve at 999 9 mL h initial 1 4 1 3 0 9 1 0 1 1 1 1 1 2 1 8 1 3 1 5 1 2 15 10 5 0 5 10 15 0 5 10 15 20 25 30 Observation Interval min Flow Rate Error Pressure Mode Trumpet Curve...

Page 43: ...4 8 3 7 11 5 0 6 23 4 2 6 32 6 13 3 1 2 15 10 5 0 5 10 15 0 5 10 15 20 25 30 Observation Interval min Flow Rate Error Resistance Mode Trumpet Curve at 0 1 mL h 48 hr 22 9 22 0 46 4 17 2 48 9 6 7 109...

Page 44: ...erval min Flow Rate Error Resistance Mode Trumpet Curve at 999 9 mL h 24 hr 3 5 3 4 2 7 2 9 3 1 3 1 3 9 3 3 3 7 3 2 3 3 15 10 5 0 5 10 15 0 5 10 15 20 25 30 Observation Interval min Flow Rate Error Hi...

Page 45: ...mark CareFusion Via Ticino 4 50019 Sesto Fiorentino Firenze Italia CareFusion Hammarbacken 4B 191 46 Sollentuna Sverige Tel 32 2 267 38 99 Tlf 45 70 20 30 74 T l 39 055 30 33 93 00 Tel 46 8 544 43 200...

Page 46: ...on This Class A b digital apparatus meets all requirements of the Canadian Interference Causing Equipment Regulation Tables The Pump is intended for use in the electromagnetic environments specified i...

Page 47: ...y should be that of a typical commercial or hospital environment IEC EN 61000 4 8 Power Frequency Magnetic Field 50 60 Hz 3 A m 400 A m 50 Hz Note 2 Power frequency magnetic fields should be at levels...

Page 48: ...arked with the following symbol Note 1 At 80 MHz and 800 MHz the higher frequency range applies Note 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absor...

Page 49: ...2 00 23 00 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be determined using the equation applicable to the frequency of the t...

Page 50: ...ument contains proprietary information of CareFusion Corporation or one of its subsidiaries and its receipt or possession does not convey any rights to reproduce its contents or to manufacture or sell...

Reviews: