1000DF00480 Issue 2 45/48
Compliance
This system complies with part 18 of the FCC Rules. Operation is subject to the following 2 conditions:
•
This system may not cause harmful interference.
•
This system must accept any interference received, including interference that may cause undesired operation.
The digital apparatus does not exceed the Class B limits for radio noise emissions from digital apparatus set out in the radio
interference regulations of the Canadian Department of Communications (DOC).
This system has been tested and found to comply with either the limits for a Class B digital pump (without Model 180 Flow
Sensor), or as a Class A digital pump (with Model 180 Flow Sensor), pursuant to Part 18 of the FCC Rules. These limits are designed
to provide reasonable protection against harmful interference when the system is operated in a commercial environment. This
system generates, uses, and can radiate radio frequency energy. If it is not installed and used in accordance with the applicable
directions for use, it may cause harmful interference to radio communications. Operation of this system in a residential area is
likely to cause harmful interference, in which case the user will be required to correct the interference at their own expense.
The authority to operate this system is conditioned by the requirement that no modifications will be made to the system unless
the changes or modifications are expressly approved by CareFusion.
This Class A/B digital apparatus meets all requirements of the Canadian Interference-Causing Equipment Regulation.
Tables:
The Pump is intended for use in the electromagnetic environments specified in the following tables.
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The pump is intended for use in the electromagnetic environment specified below.
The customer or the user of the pump should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment – Guidance
CISPR 11
RF Emissions
Group 1
Class A
The pump is suitable for use in all establishments, including domestic
establishments and those directly connected to a public low-voltage power
supply network that supplies buildings used for domestic purposes, provided the
following Caution is heeded:.
Caution:
The pump is intended for use under the supervision of healthcare professionals
only. This is a CISPR 11 Class A when the Model 180 (Flow Sensor) accessory
is used and a CISPR 11 Class B when the Model 180 is not used. In a domestic
environment, this system may cause radio interference. Reorienting, relocating or
shielding the system, or filtering the connection to the public mains network, are
examples of steps that can be taken to reduce or eliminate interference.
CISPR 11
RF Emissions
Group 1
Class B
The pump is suitable for use in all establishments, including domestic
establishments and those directly connected to a public low-voltage power
supply network that supplies buildings used for domestic purposes.
IEC/EN 61000-3-2
Harmonic Emissions
Class A
IEC/EN 61000-3-3
Voltage Fluctuations,
Flicker Emissions
Complies