![CareFusion Alaris SE 7130 Directions For Use Manual Download Page 43](http://html1.mh-extra.com/html/carefusion/alaris-se-7130/alaris-se-7130_directions-for-use-manual_2557175043.webp)
1000DF00480 Issue 2 42/48
Trumpet and Start-Up Curves (continued)
Resistance Mode
Resistance Mode Start-up at 0.1 mL/h (initial)
-0.02
0.00
0.02
0.04
0.06
0.08
0.10
0.12
0.14
0.16
0.18
0.20
0
20
40
60
80
100
120
Time (min)
(mL/h
)
Resistance Mode Start-up at 1 mL/h (initial)
-0.2
0.0
0.2
0.4
0.6
0.8
1.0
1.2
1.4
1.6
1.8
2.0
0
20
40
60
80
100
120
Time (min)
(mL/h
)
Resistance Mode Trumpet Curve at 0.1 mL/h (initial)
-22.7
-19.0
43.0
18.8
-49.2
-6.3
-104.7
14.6
153.4
73.0
1.8
-100
-80
-60
-40
-20
0
20
40
60
80
100
0
5
10
15
20
25
30
Observation Interval (min)
Flow
Rate
Erro
r(
%)
Resistance Mode Trumpet Curve at 1 mL/h (initial)
-1.4
-1.6
4.8
3.7
-11.5
-0.6
-23.4
2.6
32.6
13.3
1.2
-15
-10
-5
0
5
10
15
0
5
10
15
20
25
30
Observation Interval (min)
Flow
Rate
Erro
r(
%)
Resistance Mode Trumpet Curve at 0.1 mL/h (48 hr)
-22.9
-22.0
46.4
17.2
-48.9
-6.7
-109.5
13.3
173.6
74.6
-1.5
-100
-80
-60
-40
-20
0
20
40
60
80
100
0
5
10
15
20
25
30
Observation Interval (min)
Flow
Rate
Erro
r(
%)
Resistance Mode Trumpet Curve at 1 mL/h (48 hr)
-0.6
-1.0
5.9
4.2
-11.7
0.1
-23.5
3.2
33.1
14.2
1.7
-15
-10
-5
0
5
10
15
0
5
10
15
20
25
30
Observation Interval (min)
Flow
Rate
Erro
r(
%)
Resistance Mode Start-up at 25 mL/h (initial)
-5
0
5
10
15
20
25
30
35
40
45
50
0
20
40
60
80
100
120
Time (min)
(mL/h
)
Resistance Mode Start-up at 999.9 mL/h (initial)
-200
0
200
400
600
800
1000
1200
1400
1600
1800
2000
0
20
40
60
80
100
120
Time (min)
(mL/h
)
Legend:
Maximum rate error
y
Overall rate error
u
Minimum rate error