1000DF00480 Issue 2 27/48
Alarms, Alerts and Prompts (continued)
Prompt
Meaning
Response
Air In Line
Air detector has detected air prior to starting
infusion or is in poor contact with set.
Press
continue
soft key to allow infusion to
continue. An alarm occurs if air detector detects
an air bubble larger than configured threshold.
Verify set is loaded correctly. Prime and reload
set or remove air. Reshape tubing to ensure
optimum contact with sensor.
Ensure air-in-line sensors are thoroughly
cleaned.
Dose Out of Range
Calculated dose is outside allowable range.
Verify and re-enter settings.
Entry Invalid
An invalid value was entered during programming.
Press
CLEAR
or
0
key to clear entry. Enter
appropriate value.
Instrument Self-Check Is Due
Please Eject the Set
Pump/channel has not performed self-check within
programmed interval.
If set is loaded: Eject set, wait 5 seconds and
then reload set.
If no set is loaded: Load set, wait 1 minute and
then eject set. Wait 5 seconds and then reload
set.
Invalid Entry Rate Out of
Range
Pump has calculated a rate less than 0.1ml/h.
Verify and re-enter settings.
Latch Open
Latch is open (prior to starting an infusion).
Close latch
fully
to left.
Maintenance Reminder
Periodic maintenance interval has elapsed.
Maintenance Reminder feature is on.
Notify Biomedical Engineering department. If
desired, press
continue
soft key to temporarily
bypass reminder.
Max Rate = XXX.X ml/h
(
XXX.X represents maximum
flow rate configured for pump
or profile.
)
An attempt was made to enter a rate greater than
maximum configured rate or pump has calculated
a rate greater than maximum configured rate.
Default maximum rate is 999.9 ml/h in primary
mode, 600 ml/h in secondary mode.
Verify and re-enter settings.
New Baseline Set
A new Manual Pressure Baseline was successfully
set.
Manual Pressure Baseline feature is on.
Baseline remains set until a new manual baseline
is set, pump is turned off or latch is opened.
Occlusion Downstream
A very high pressure exists in infusion set while
baseline is being set.
Pressure Baseline feature is on.
Remove source of high pressure and repeat
setting of pressure baseline.
Ok Entry
Attempt was made to go to another page before
pressing
ok
soft key.
Verify all parameters are correct and press
ok
soft key.
Panel Locked
A key was pressed.
Panel lock feature is on.
Turn panel lock off to access panel controls.
Panel lock key is located behind handle.
Place on Hold to Change
A key
was pressed during KVO.
Channel must be on hold to make changes.
Place on Hold to Set Pressure
Baseline
SET PRESSURE BASELINE function was selected
while running.
Pressure Baseline feature is on.
Place pump on hold before performing manual
SET PRESSURE BASELINE operation.
Press and Hold Key to Turn
Off
POWER
was pressed.
Press and hold
POWER
until display turns off.
Pressure Limit XXX mmHg
(
XXX represents configured
maximum pressure
)
An elevated pressure was present in fluid path
when pressure baseline was established. This may
reduce maximum available pressure range.
Reload infusion set and verify no obstruction
exists which could cause excess pressure.
• If
Pressure Baseline feature is on, repeat manual
setting of pressure baseline.
otherwise
• Restarting infusion automatically sets pressure
baseline.
Pressure Limit Must Be Less
Than or Equal to XXX mmHg
(
XXX represents configured
maximum pressure
)
Attempt was made to increase pressure alarm
limit to a level higher than configured maximum
pressure.
Choose a pressure alarm limit that is less than, or
equal to, configured maximum pressure.
Prompts