![CareFusion Alaris SE 7130 Directions For Use Manual Download Page 11](http://html1.mh-extra.com/html/carefusion/alaris-se-7130/alaris-se-7130_directions-for-use-manual_2557175011.webp)
1000DF00480 Issue 2 10/48
Getting Started
Initial Set Up
1. Check that the pump is complete, undamaged and that the voltage rating specified on the label is compatible with your
AC power supply.
2. Items supplied are :
•
Alaris® SE Pump
•
Directions For Use (CD)
•
AC Power Cable (as requested)
•
Protective Packaging
3. Connect the pump to the AC power supply for at least 24 hours to ensure that the internal battery is charged (verify that
the AC Mains indicator is lit).
Pole Clamp Installation
A
Before operating the pump read this Directions For Use (DFU) manual carefully.
A
Maximum battery capacity, as well as gauge accuracy, is reached after several charge/discharge/
recharge cycles, in the refresh process. CareFusion recommends that the battery be fully charged/
discharged/recharged, using the refresh cycle, before placing the pump in use.
The pump will automatically operate from its internal battery if the pump is switched on without being
connected to the power supply.
Should the pump fail to perform correctly, replace in its original protective packaging, where possible
and contact a qualified service engineer for investigation.
The uniquely designed pole clamp adapts to a wide variety of surfaces (such as,
poles, bed rails) to provide greater versatility and to simplify transports.
It features:
•
360º rotation in 90º increments
•
ergonomically designed knob
•
accommodates diameters from 15 to 35 millimetres
A
B
A
B
A
B
A
B
Changing Pole Clamp Orientation
1. Press and hold rotation lever.
2. Reposition clamp.
3. Release lever at desired position.
A
B
A
B
Lever
A
B
A
B
A
•
The illustrated pole clamp knob may not reflect the knob in use on the pump.
•
When using multiple pumps, care should be taken to evenly distribute the pumps to ensure
stability.
•
To ensure proper occlusion detection, do not operate the pump tilted back more than 45º from the
upright position.