1000DF00480 Issue 2 24/48
Monitoring Options (continued)
Options (continued)
Pressure Trend Graphs
In Pressure Monitoring mode, a trend graph displays monitored pressure over time. Trend graphs of 15 minutes, 1 hour, 4 hours
and 12 hours are available during normal operation when enabled through the hospital data set.
Downstream Occlusions, which occur in Pressure or Resistance modes, are indicated by a tick mark (
y
) at the top of the trend
screen.
To view graph:
1. Dual channel pumps: Select desired channel, as necessary. Trend graph is not available while split screen is displayed.
2. Press
OPTIONS
key.
3. Press
Pressure Trend
soft key.
4. To change graph time frame, press
time
soft key.
A
•
A solid horizontal line represents current pressure alarm limit level.
•
Gaps in graph may indicate non-infusing conditions; such as, turned off, on hold, in alarm.
•
If channel has been placed in Resistance Monitoring mode for some portion of a trend graph
window, pressure data is not available and zero values are plotted.
•
A tick mark (
y
) at top of graph indicates an occlusion.
To clear graph:
1. To clear graphed data, press
clear
soft key.
2. Press
ok
soft key.
To return to normal operation:
1. Press
return
soft key.
SmartSite® Needle-Free System Instructions
DIRECTIONS - Use Aseptic Technique
1. Prior to every access, swab top of Needle-Free Valve port with 70% Isopropyl alcohol (1-2
seconds) and allow to dry (approximately 30 seconds).
NOTE: Dry time is dependent on temperature, humidity, ventilation of the area.
2. Prime valve port. If applicable, attach syringe to Needle-Free Valve port and aspirate minuscule
air bubbles.
3.
Replace every 72 hours for stand alone valves. However, if the valve is part of the set, then the set
change interval is as per the complete set or 100 activations which ever occurs first. For infusions
of blood, blood products or lipid emulsions replace infusion set every 24 hours.
NOTE: During use of Needle-Free Valve port, fluid may be observed between the housing and blue
piston. This fluid does not enter the fluid path and requires no action.
For product questions or needle-free valve educational materials contact your Cardinal
Health, Alaris® Products representative. The Center for Disease Control, Intravenous
Nurses Society (USA) and other organizations publish guidelines useful in developing
facility guidelines. Consult facility protocols.
A
Precautions:
Discard if packaging is not intact or protector caps are unattached.
Needle-Free Valve contraindicated for blunt cannula system.
SmartSite® Needle-Free Valve is designed to permit safe gravity flow and automated flow, injection and aspiration of fluids without the use of
needles by utilising Luer lock and luer slip connectors.