![CareFusion Alaris SE 7130 Directions For Use Manual Download Page 19](http://html1.mh-extra.com/html/carefusion/alaris-se-7130/alaris-se-7130_directions-for-use-manual_2557175019.webp)
1000DF00480 Issue 2 18/48
Multi-Dose
This feature allows 1 to 24 infusions to be preprogrammed with the same rate and volume, to be delivered at equally spaced
intervals, over a period of up to 24 hours. It also offers a delayed start option up to 8 hours and a Dose Complete Alert Option.
These features can be turned on or off.
This program requires another infusing line to keep the vein open between programmed doses since there is no KVO infusion
between doses or following program completion.
1. Select desired channel, as necessary. Channel must be on hold in primary mode.
2. Press
OPTIONS
key.
3. Press
Multi-Dose
soft key.
4. Press
Enter New Program
soft key.
5. To enter infusion rate, use numeric keypad. Press
ENTER
key.
6. To enter VTBI/Dose, use numeric keypad. Press
ENTER
key.
7. To enter number of doses, use numeric keypad. Press
ENTER
key.
8. To enter dose frequency (time interval from start of one dose until start of next), use numeric keypad. Press
ENTER
key.
9. Verify that all parameters are correct then press
ok
soft key.
10. If Dose Complete Alert Option is enabled,
DOSE COMPLETE ALERT OPTION
page appears. To select
On
or
Off
, use soft
keys.
11. To continue programming, press
ok
soft key.
12. To start first dose immediately, press
ok
soft key. Then press
run
soft key or
RUN/HOLD
key to start infusion.
13. To delay start of first dose, enter time (hours and minutes) using numeric keypad and
ENTER
key. Then to advance to timer
hold page, press
start timer
soft key.
Changing Time Interval Until Next Dose
1. Press
stop timer
soft key.
2. To select a value for editing, press relevant soft key.
3. To enter new value, use numeric keypad. Press
ENTER
key.
4. When editing is complete, press
start timer
soft key.
Resuming an Interrupted Multi-Dose
1. Press
yes
soft key on the
Return To Multi-Dose?
page.
2. To accept set-up parameters, press
Review/Resume
soft key.
3. To continue, press
ok
soft key.
4. If infusion was in progress when interrupted press
run
soft key or
RUN/HOLD
key to resume infusion. If infusion was NOT
in progress when interrupted, edit time to delivery of next dose, as necessary, then press
start timer
soft key.
Quitting Multi-Dose
The channel must be on hold or the last dose complete.
1. Press
menu
soft key.
2. To return to primary set-up page, press
Quit Program
soft key.