Alarms, Alerts and Prompts (continued)
Prompts (continued)
Prompt
Meaning
Response
Pressure Unstable Cannot
Set Baseline
Excessive variation in pressure due to motion, flow
from other pumps or blood pressure prevents
accurate setting of pressure baseline.
Pressure Baseline feature is on.
Reduce or temporarily remove sources of
variation while performing manual baseline
setting operation.
Program Lost Re-Enter
Settings
Pump detected a memory or power failure. Existing
operating parameters were erased.
Press
continue
soft key and re-enter all infusion
settings. Configurable options are not affected.
Rate Out of Range
Pump has calculated a rate less than 0.1 ml/h.
Verify and re-enter settings.
Set Must Be Loaded
Flow Regulator segment is not loaded in selected
channel during a manual pressure baseline setting
operation.
Pressure Baseline feature is on.
Load Flow Regulator segment in selected
channel. Repeat manual pressure baseline
setting.
Set Out
Flow Regulator segment is not installed correctly.
Reinstall Flow Regulator segment.
Set Pressure Baseline
Set Pressure Baseline was selected in options mode.
Press
ok
soft key to set Pressure Baseline or press
return
soft key to go to exit.
Set Pri VTBI
A primary VTBI was not programmed.
Enter a primary VTBI.
Set Pri VTBI > Loading Dose
VTBI
Loading Dose VTBI entered is greater than primary
VTBI.
Raise primary VTBI or lower Loading Dose VTBI,
as appropriate.
Stop Timer to Change
An invalid key was pressed while timer was running
in Multi-Dose program.
Wait several seconds for popup to finish. Press
stop timer
soft key to make changes.
Time Out of Range
Programmed step time exceeds 24 hours and 59
minutes, or is less than 1 minute.
Verify and re-enter settings.
Soft Limit
Programmed parameter is outside Soft Limit for
that care area.
If it is clinically appropriate and necessary to
override Soft Limit, press
yes
soft key.
Otherwise press
no
soft key and enter new value
within limit.
Impotant
: Prior to overriding a Soft Limit
prompt, confirm the infusion parameters are
correct.
Hard Limit
Programmed parameter is outside Hard Limit for
that care area.
Use numeric keys to enter a new value, then
press
ENTER
key.
Invalid Keypress During Programming:
The following Prompts may be seen if an invalid key is pressed during programming: Both A and B not Running, Channel Not On,
Complete or OK Setup, No Numeric Entries, Select Channel.
Invalid Keypress During Infusion:
During an infusion, if an invalid key is pressed, the following prompts may be seen: Dose Rate Running, Load Dose Running,
Multi-Dose Running, Multi-Step Running, Pri Running, Sec Running, or Timer Running (Multi-Dose program).
Resuming Previous Programming:
When a pump has been powered off then on again previous parameters may be preserved if Current Profile is accepted and
New Patient? – No
is selected. The following prompts may be seen: Return to Dose Rate?, Return to Loading Dose?, Return to
Multi-Dose?, Return to Multi-Step?, or Return to Secondary?.