TROUBLESHOOTING
6-3
Table 6-2.
Malfunction Codes
Refer to the PC Unit / Pump Module Technical Service Manual for information on viewing the
Malfunction Log.
Alaris
®
EtCO
2
Module (8300 Series)
Technical Service Manual
Malfunction
Code
Subsystem
Explanation
Remedy
500
Main Logic Board
510
Keypad Decoder
511
Keypad Decoder Comm Error
530
Keypad Error
550
IUI Error
560
Power Supply Board Error
570
EtCO
2
Board Error
571
EtCO
2
Board Comm Error
6200
EtCO
2
Board Error
Failed continuous built-in
tests.
Cycle power. If error repeats,
replace EtCO
2
Board.
6210
EtCO
2
Board Error
Communications error.
Cycle power. If error repeats,
replace EtCO
2
Board.
6220
EtCO
2
Board Error
Configuration error.
Cycle power. If error repeats,
replace EtCO
2
Board.
6230
EtCO
2
Board Error
Received value out of
range.
Replace Microstream
Disposable device. If error
repeats, replace EtCO
2
Board.
6240
EtCO
2
Board Error
Missing sensor status.
Replace Microstream
Disposable device. If error
repeats, replace EtCO
2
Board.
NOTE:
Refer to Troubleshooting information in the PC Unit / Pump Module Technical Service Manual for more
detailed descriptions of codes.
Summary of Contents for Alaris 8300 Series
Page 7: ...1 GENERAL INFORMATION...
Page 8: ......
Page 11: ...2 CHECKOUT CONFIGURATION...
Page 12: ......
Page 21: ...3 PREVENTIVE MAINTENANCE...
Page 22: ......
Page 25: ...4 PRINCIPLES OF OPERATION...
Page 26: ......
Page 29: ...5 CORRECTIVE MAINTENANCE...
Page 30: ......
Page 47: ...6 TROUBLESHOOTING...
Page 48: ......
Page 53: ...7 ILLUSTRATED PARTS BREAKDOWN...
Page 54: ......
Page 67: ......