Chapter 4 — PRINCIPLES OF OPERATION
4-1
Alaris
®
EtCO
2
Module (8300 Series)
Technical Service Manual
4.1
INTRODUCTION
This chapter describes the principles of
operation for the EtCO
2
Module and its
major subsystems.
Refer to the PC Unit/Pump Module
Technical Service Manual for Point-of-Care
Unit information.
4.2
GENERAL INFORMATION
The EtCO
2
Module uses Oridion’s patented
Microstream
®
non-dispersive infrared (NDIR)
spectroscopy to continuously measure the
amount of CO
2
during every breath, the
amount of CO
2
present at the end of
exhalation (EtCO
2
) and during inhalation
(FiCO
2
), and the Respiratory Rate. The
EtCO
2
Module is a side stream capnograph.
The Microstream
®
Disposables deliver a
sample of the inhaled and exhaled gases
from the ventilator disposable or directly
from the patient (via an oral/nasal cannula)
into the monitor for CO
2
measurement.
Moisture and patient secretions are
extracted from the sample by the
Microstream
®
inline filter while maintaining
the shape of the CO
2
waveform.
The 50 mL/min sampling flow rate reduces
liquid and secretion accumulation,
decreasing the risk of obstruction in the
sample pathway in humid ICU
environments. The small sample size
required by Microstream
®
eliminates the
need for water traps and prevents excess
fluid accumulation.
The EtCO
2
Module draws a gas sample
through a microsample cell (15 microliters).
This extremely small volume is quickly
flushed, allowing for a rise time of
Summary of Contents for Alaris 8300 Series
Page 7: ...1 GENERAL INFORMATION...
Page 8: ......
Page 11: ...2 CHECKOUT CONFIGURATION...
Page 12: ......
Page 21: ...3 PREVENTIVE MAINTENANCE...
Page 22: ......
Page 25: ...4 PRINCIPLES OF OPERATION...
Page 26: ......
Page 29: ...5 CORRECTIVE MAINTENANCE...
Page 30: ......
Page 47: ...6 TROUBLESHOOTING...
Page 48: ......
Page 53: ...7 ILLUSTRATED PARTS BREAKDOWN...
Page 54: ......
Page 67: ......