Follow these procedures in order for the
most efficient disassembly of EtCO
2
Module
subassemblies. To reassemble the EtCO
2
Module, reverse the disassembly steps.
NOTE:
Perform all maintenance on an
antistatic surface, preferably a grounded
antistatic mat.
Before adhering gaskets and labels to the
instrument, clean the surface with a cotton
swab or soft cloth lightly dampened with
70%
Isopropyl Alcohol.
5-2
Alaris
®
EtCO
2
Module (8300 Series)
Technical Service Manual
CORRECTIVE MAINTENANCE
5.2
DISASSEMBLY/REASSEMBLY
Disconnect the EtCO
2
Module from the
Alaris
®
System Before attempting
disassembly.
CAUTION
To avoid the risk of electrical hazard or
damage to the instrument circuitry, do not
spray fluids directly onto the instrument or
allow fluids to enter the instrument.
CAUTION
Table 5-1.
Required Materials, Supplies and Tools
NOTE:
Contact/source information is subject to change.
•
Loctite
242
(
http://www.loctite.com
)
•
Silicone grease, Dow Corning Molykote
33
, or equivalent (
http://www.dowcorning.com
)
•
#1
Phillips screwdriver
•
#2
Phillips screwdriver
•
Small diagonal cutters
•
Lint-free cloth (such as, Kimwipes or lint-free tissue)
Summary of Contents for Alaris 8300 Series
Page 7: ...1 GENERAL INFORMATION...
Page 8: ......
Page 11: ...2 CHECKOUT CONFIGURATION...
Page 12: ......
Page 21: ...3 PREVENTIVE MAINTENANCE...
Page 22: ......
Page 25: ...4 PRINCIPLES OF OPERATION...
Page 26: ......
Page 29: ...5 CORRECTIVE MAINTENANCE...
Page 30: ......
Page 47: ...6 TROUBLESHOOTING...
Page 48: ......
Page 53: ...7 ILLUSTRATED PARTS BREAKDOWN...
Page 54: ......
Page 67: ......