Operator´s Manual
1000-095-071-EN Rev. 1.1
OMNERA®
2020-09-14
Page 91
1
2
3
Fig. 3-39 Wireless foot control.
1. Z movement down
2. Z movement up
3. Release pedal:
Unlocks wallstand detector brake. Manual movement is possible.
Unlocks table top. Manual movement is possible, table top is floating.
3.7.4.1 How to Manoeuver
1. Press pedal to move downward.
2. Press pedal to move upward.
3. Press the pedal to release the brakes. On activation, the table top or the wallstand detec-
tor holder, can be moved manually.
When the release pedal is activated, the collimator light will be lit.
Note!
The foot control must not be used outside the examination room.
The device has no applied parts and should not be accessible to patients.
3.7.4.2 Battery
The yellow battery indication LED will begin to flash once every two seconds, when the re-
maining battery capacity is approximately 1 week of constant use, or 168 hours. It will then
change to 2 flashes per second when the capacity has been reduced to approximately 2
days, or 48 hours.