OMNERA®
1000-095-071-EN Rev. 1.1
Operator´s Manual
Page 124
2020-09-14
a)
b)
Fig. 4-27 a) Left position and b) Right position.
Activate the mode by pressing the servo button.
• The servo mode indication light will flash until the left and right positions are specified.
• The OTC will automatically move to its programmed position.
• The system will beep when position is reached.
Invite the patient and position the patient on the table.
Move the focus point to the middle of the planned composite image. This could be done ei-
ther by moving the tabletop or by moving the column in X (or Y) direction.
Rotate the x-ray tube in order for the right edge of the collimator light field to indicate the right
limit for the composite image.
Press the button (F1) to set the right limit, see Fig. 4-28 .
• The button turns green to indicate that the limit is set.
F2
F1
Fig. 4-28