![Canon Omnera 500A Operator'S Manual Download Page 96](http://html.mh-extra.com/html/canon/omnera-500a/omnera-500a_operators-manual_61481096.webp)
OMNERA®
1000-095-071-EN Rev. 1.1
Operator´s Manual
Page 88
2020-09-14
3.6 Servo Button
When the servo button is activated, the OTC and Wall stand detector (depending on system
setup) automatically starts positioning (excluding
Free mode
). If the servo button is released,
the movement stops and a manual movement of the stand is possible.
When a new protocol is chosen or if the OTC is manually moved from the position, the OTC
automatically starts moving to a predefined position when the servo button is activated.
The servo button is available on the external servo button module, the Wall stand hand con-
trol, and on the remote control (option). The system status is indicated by indication light on
the external servo module and at the OTC.
3.6.1 External Servo Button Module
The external servo button module holds a servo button and an emergency stop. The servo
button is equipped with an indication light showing the system status. See Fig. 3-35.
20
30-01
1
2
Fig. 3-35 External servo button module
1. Emergency stop
2. Servo button and Indication light
3.6.2 Indication Light
WARNING!
Due to the squeezing hazard, motorized movements are only allowed if patient and
System are observed by personnel.
The Indicator light has 3 different modes:
•
Fixed light
- The System is in position and ready for exposure. Supervision of the patient
and System is required.
•
Flashing light
- Attention. The System is performing an action, for instance manoeuvering
to start position or waiting for action.
•
No light
- the chosen mode is not activated. No light is shown in
Free mode
.
See also 3.2.14
3.6.3 DAP
If a DAP meter is included the System, the Dose Area Product will be presented in the Imag-
ing system.
Checks and settings can be done by the service software, see
‘Installation and Service Man-
ual’
, Chapter
5, Installation.